EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Fully Automated Assessment of Cardiac Chamber Volumes and Myocardial Mass on Non-Contrast Chest CT with a Deep Learning Model: Validation Against Cardiac MR

Ramona Schmitt iD, Christopher L Schlett iD, Jonathan I Sperl iD, Saikiran Rapaka, Athira J Jacob iD, Manuel Hein iD, Muhammad Taha Hagar iD, Philipp Ruile, Dirk Westermann, Martin Soschynski iD, Fabian Bamberg, Christopher Schuppert iD

Diagnostics · 21.12.2024 · 14 (24) · 2884

DOI: 10.3390/diagnostics14242884 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 21.12.2024. Verfasst von Ramona Schmitt und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology and Angiology, University Heart Center Freiburg-Bad Krozingen, Medical Center-University of Freiburg, Faculty of Medicine, U…. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

To validate the automated quantification of cardiac chamber volumes and myocardial mass on non-contrast chest CT using cardiac MR (CMR) as a reference.

Methods

We retrospectively included 53 consecutive patients who received non-contrast chest CT and CMR within three weeks. A deep learning model created cardiac segmentations on axial soft-tissue reconstructions from CT, covering all four cardiac chambers and the left ventricular myocardium. Segmentations on CMR cine short-axis and long-axis images served as a reference. Standard estimates of diagnostic accuracy were calculated for ventricular volumes at end-diastole and end-systole (LVEDV, LVESV, RVEDV, RVESV), left ventricular mass (LVM), and atrial volumes (LA, RA) at ventricular end-diastole. A qualitative assessment noted segmentation issues.

Results

The deep learning model generated CT measurements for 52 of the 53 patients (98%). Based on CMR measurements, the average LVEDV was 166 ± 64 mL, RVEDV was 144 ± 51 mL, and LVM was 115 ± 39 g. The CT measurements correlated well with CMR measurements for LVEDV, LVESV, and LVM (ICC = 0.85, ICC = 0.84, and ICC = 0.91; all p < 0.001) and RVEDV and RVESV (ICC = 0.79 and ICC= 0.78; both p < 0.001), and moderately well with LA and RA (ICC = 0.74 and ICC = 0.61; both p < 0.001). Absolute agreements likewise favored LVEDV, LVM, and RVEDV. ECG-gating did not relevantly influence the results. The CT results correctly identified 7/15 LV and 1/1 RV as dilated (one and six false positives, respectively). Major qualitative issues were found in three cases (6%).

Conclusions

Automated cardiac chamber volume and myocardial mass quantification on non-contrast chest CT produced viable measurements in this retrospective sample. Relevance Statement: An automated cardiac assessment on non-contrast chest CT provides quantitative morphological data on the heart, enabling a preliminary organ evaluation that aids in incidentally identifying at-risk patients who may benefit from a more targeted diagnostic workup.

Schlagwörter

Stichwörter der Autor:innen

cardiovascular diseaseschest CTdeep learningincidental findingsnon-contrast CT

Autor:innen und Institutionen

  • Department of Cardiology and Angiology, University Heart Center Freiburg-Bad Krozingen, Medical Center-University of Freiburg, Faculty of Medicine, University of Freiburg, Südring 15, 79189 Bad Krozingen, Germany.
  • Department of Diagnostic and Interventional Radiology, Medical Center-University of Freiburg, Faculty of Medicine, University of Freiburg, Hugstetter Str. 55, 79106 Freiburg im Breisgau, Germany.
  • Siemens Healthineers, Siemensstr. 1, 91301 Forchheim, Germany.
  • Siemens Healthineers, 755 College Rd E, Princeton, NJ 08540, USA.

ORCID-Profile: Ramona Schmitt, Christopher L Schlett, Jonathan I Sperl, Athira J Jacob, Manuel Hein, Muhammad Taha Hagar, Martin Soschynski, Christopher Schuppert

Zitieren

APA

Schmitt, R., Schlett, C. L., Sperl, J. I., Rapaka, S., Jacob, A. J., Hein, M., Hagar, M. T., Ruile, P., Westermann, D., Soschynski, M., Bamberg, F., & Schuppert, C. (2024). Fully Automated Assessment of Cardiac Chamber Volumes and Myocardial Mass on Non-Contrast Chest CT with a Deep Learning Model: Validation Against Cardiac MR. Diagnostics, 14(24), 2884. https://doi.org/10.3390/diagnostics14242884

Vancouver

Schmitt R, Schlett CL, Sperl JI, Rapaka S, Jacob AJ, Hein M, et al. Fully Automated Assessment of Cardiac Chamber Volumes and Myocardial Mass on Non-Contrast Chest CT with a Deep Learning Model: Validation Against Cardiac MR. Diagnostics. 2024;14(24):2884. doi:10.3390/diagnostics14242884

BibTeX
@article{Schmitt2024,
  author = {Schmitt, Ramona and Schlett, Christopher L and Sperl, Jonathan I and Rapaka, Saikiran and Jacob, Athira J and Hein, Manuel and Hagar, Muhammad Taha and Ruile, Philipp and Westermann, Dirk and Soschynski, Martin and Bamberg, Fabian and Schuppert, Christopher},
  title = {Fully Automated Assessment of Cardiac Chamber Volumes and Myocardial Mass on Non-Contrast Chest CT with a Deep Learning Model: Validation Against Cardiac MR},
  journal = {Diagnostics},
  year = {2024},
  volume = {14},
  number = {24},
  pages = {2884},
  doi = {10.3390/diagnostics14242884}
}
Kostenlos starten →