Comparison of Bone Evaluation and Metal Artifact between Photon-Counting CT and Five Energy-Integrating-Detector CT under Standardized Conditions Using Cadaveric Forearms

Takeshi Fukuda iD, Takenori Yonenaga, Ryo Akao, Tohru Hashimoto, Kazuhiro Maeda, Tomokazu Shoji, Shoichi Shioda, Yu Ishizaka, Hiroya Ojiri iD

Diagnostics · 06.02.2024 · 14 (4) · 350

DOI: 10.3390/diagnostics14040350 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 06.02.2024. Verfasst von Takeshi Fukuda und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, The Jikei University School of Medicine, 3-25-8 Nishi-Shimbashi, Minato-ku, Tokyo 105-8461, Japan. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

To compare the potential of various bone evaluations by considering photon-counting CT (PCCT) and multiple energy-integrating-detector CT (EIDCT), including three dual-energy CT (DECT) scanners with standardized various parameters in both standard resolution (STD) and ultra-high-resolution (UHR) modes.

Methods

Four cadaveric forearms were scanned using PCCT and five EIDCTs, by applying STD and UHR modes. Visibility of bone architecture, image quality, and a non-displaced fracture were subjectively scored against a reference EIDCT image by using a five-point scale. Image noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were also compared. To assess metal artifacts, a forearm with radial plate fixation was scanned by with and without Tin filter (Sn+ and Sn−), and virtual monoenergetic image (VMI) at 120 keV was created. Regarding Sn+ and VMI, images were only obtained from the technically available scanners. Subjective scores and the areas of streak artifacts were compared.

Results

PCCT demonstrated significantly lower noise (p < 0.001) and higher bone SNR and CNR (p < 0.001) than all EIDCTs in both resolution modes. However, there was no significant difference between PCCT and EIDCTs in almost all subjective scores, regardless of scan modes, except for image quality where a significant difference was observed, compared to several EIDCTs. Metal artifact analysis revealed PCCT had larger artifact in Sn− and Sn+ (p < 0.001), but fewer in VMIs than three DECTs (p < 0.001 or 0.001).

Conclusions

Under standardized conditions, while PCCT had almost no subjective superiority in visualizing bone structures and fracture line when compared to EIDCTs, it outperformed in quantitative analysis related to image quality, especially in lower noise and higher tissue contrast. When using PCCT to assess cases with metal implants, it may be recommended to use VMIs to minimize the possible tendency for artifact to be pronounced.

Schlagwörter

Stichwörter der Autor:innen

boneenergy-integrating detectorfracturemetal artifactphoton-counting detector

Autor:innen und Institutionen

  • Department of Radiology, The Jikei University School of Medicine, 3-25-8 Nishi-Shimbashi, Minato-ku, Tokyo 105-8461, Japan.
  • Department of Anatomy, The Jikei University School of Medicine, 3-25-8 Nishi-Shimbashi, Minato-ku, Tokyo 105-8461, Japan.
  • Department of Orthopaedic Surgery, The Jikei University School of Medicine, 3-25-8 Nishi-Shimbashi, Minato-ku, Tokyo 105-8461, Japan.
  • Department of Radiology, Tha Jikei University Katsushika Medical Center, 6-41-2 Aoto, Katsushika-ku, Tokyo 125-8506, Japan.
  • Medicalscanning Tokyo, 3-1-17 Nihonbashi, Chuo-ku, Tokyo 103-0027, Japan.

ORCID-Profile: Takeshi Fukuda, Hiroya Ojiri

Zitieren

APA

Fukuda, T., Yonenaga, T., Akao, R., Hashimoto, T., Maeda, K., Shoji, T., Shioda, S., Ishizaka, Y., & Ojiri, H. (2024). Comparison of Bone Evaluation and Metal Artifact between Photon-Counting CT and Five Energy-Integrating-Detector CT under Standardized Conditions Using Cadaveric Forearms. Diagnostics, 14(4), 350. https://doi.org/10.3390/diagnostics14040350

Vancouver

Fukuda T, Yonenaga T, Akao R, Hashimoto T, Maeda K, Shoji T, et al. Comparison of Bone Evaluation and Metal Artifact between Photon-Counting CT and Five Energy-Integrating-Detector CT under Standardized Conditions Using Cadaveric Forearms. Diagnostics. 2024;14(4):350. doi:10.3390/diagnostics14040350

BibTeX
@article{Fukuda2024,
  author = {Fukuda, Takeshi and Yonenaga, Takenori and Akao, Ryo and Hashimoto, Tohru and Maeda, Kazuhiro and Shoji, Tomokazu and Shioda, Shoichi and Ishizaka, Yu and Ojiri, Hiroya},
  title = {Comparison of Bone Evaluation and Metal Artifact between Photon-Counting CT and Five Energy-Integrating-Detector CT under Standardized Conditions Using Cadaveric Forearms},
  journal = {Diagnostics},
  year = {2024},
  volume = {14},
  number = {4},
  pages = {350},
  doi = {10.3390/diagnostics14040350}
}
Kostenlos starten →