The Performance Characteristics of Handheld, Non-Piezoelectric Point-of-Care Ultrasound (POCUS) in the Emergency Department
Diagnostics · 21.12.2023 · 14 (1) · 17
DOI: 10.3390/diagnostics14010017 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 21.12.2023. Verfasst von Brandon Michael Wubben und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Emergency Medicine, University of Iowa, 200 Hawkins Drive, Iowa City, IA 52242, USA. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
The use of handheld point-of-care ultrasound (HH-POCUS) platforms is rapidly increasing, but the diagnostic performance of HH-POCUS in the emergency department (ED) has not been well-studied. For a period of one year, only a HH-POCUS system that uses a non-piezoelectric array (Butterfly iQ+™) was available for clinical POCUS examinations in our ED. We performed a retrospective observational study of patients who underwent cardiac, thoracic, renal, biliary, or lower extremity venous (DVT) examinations from November 2021–November 2022 and calculated performance characteristics of HH-POCUS relative to radiology imaging. A total of 381 HH-POCUS studies were evaluated. Cardiac image quality was significantly lower than lung (p = 0.002). Over half of the studies (213/381) had imaging available for comparison, and HH-POCUS identified 86.5% (32/37, (95%CI) 70.4–94.9) of prespecified emergent diagnoses, including acute cholecystitis, severely reduced left ventricular ejection fraction, pericardial effusion or tamponade, moderate or larger pleural effusion, pneumothorax, moderate or larger hydronephrosis, and DVT. For less emergent diagnoses, 84.3% (43/51, (95%CI) 70.9–92.5) were identified. Overall, HH-POCUS using a non-piezoelectric array showed modest real-world performance in the ED for cardiac, thoracic, renal, biliary, and DVT examinations. HH-POCUS may be inadequate to rule out some common ED diagnoses, but had good specificity for certain conditions such as pericardial effusion.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Emergency Medicine, University of Iowa, 200 Hawkins Drive, Iowa City, IA 52242, USA.
- Carver College of Medicine, University of Iowa, 375 Newton Road, Iowa City, IA 52242, USA.
ORCID-Profile: Brandon Michael Wubben, Hae In Yun
Zitieren
APA
Wubben, B. M., & Yun, H. I. (2023). The Performance Characteristics of Handheld, Non-Piezoelectric Point-of-Care Ultrasound (POCUS) in the Emergency Department. Diagnostics, 14(1), 17. https://doi.org/10.3390/diagnostics14010017
Vancouver
Wubben BM, Yun HI. The Performance Characteristics of Handheld, Non-Piezoelectric Point-of-Care Ultrasound (POCUS) in the Emergency Department. Diagnostics. 2023;14(1):17. doi:10.3390/diagnostics14010017
BibTeX
@article{Wubben2023,
author = {Wubben, Brandon Michael and Yun, Hae In},
title = {The Performance Characteristics of Handheld, Non-Piezoelectric Point-of-Care Ultrasound (POCUS) in the Emergency Department},