Image Quality and Quantitative PET Parameters of Low-Dose [18F]FDG PET in a Long Axial Field-of-View PET/CT Scanner

Eduardo Calderón iD, Fabian P Schmidt, Wenhong Lan iD, Salvador Castaneda-Vega iD, Andreas S Brendlin iD, Nils F Trautwein iD, Helmut Dittmann iD, Christian la Fougère iD, Lena Sophie Kiefer iD

Diagnostics · 18.10.2023 · 13 (20) · 3240

DOI: 10.3390/diagnostics13203240 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 18.10.2023. Verfasst von Eduardo Calderón und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nuclear Medicine and Clinical Molecular Imaging, University Hospital Tuebingen, 72076 Tuebingen, Germany. Die Arbeit wurde bisher 40-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft, Faculty of Medicine, University of Tuebingen, Faculty of Medicine, Tuebingen, Germany.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PET/CT scanners with a long axial field-of-view (LAFOV) provide increased sensitivity, enabling the adjustment of imaging parameters by reducing the injected activity or shortening the acquisition time. This study aimed to evaluate the limitations of reduced [18F]FDG activity doses on image quality, lesion detectability, and the quantification of lesion uptake in the Biograph Vision Quadra, as well as to assess the benefits of the recently introduced ultra-high sensitivity mode in a clinical setting. A number of 26 patients who underwent [18F]FDG-PET/CT (3.0 MBq/kg, 5 min scan time) were included in this analysis. The PET raw data was rebinned for shorter frame durations to simulate 5 min scans with lower activities in the high sensitivity (HS) and ultra-high sensitivity (UHS) modes. Image quality, noise, and lesion detectability (n = 82) were assessed using a 5-point Likert scale. The coefficient of variation (CoV), signal-to-noise ratio (SNR), tumor-to-background ratio (TBR), and standardized uptake values (SUV) including SUVmean, SUVmax, and SUVpeak were evaluated. Subjective image ratings were generally superior in UHS compared to the HS mode. At 0.5 MBq/kg, lesion detectability decreased to 95% (HS) and to 98% (UHS). SNR was comparable at 1.0 MBq/kg in HS (5.7 ± 0.6) and 0.5 MBq/kg in UHS (5.5 ± 0.5). With lower doses, there were negligible reductions in SUVmean and SUVpeak, whereas SUVmax increased steadily. Reducing the [18F]FDG activity to 1.0 MBq/kg (HS/UHS) in a LAFOV PET/CT provides diagnostic image quality without statistically significant changes in the uptake parameters. The UHS mode improves image quality, noise, and lesion detectability compared to the HS mode.

Schlagwörter

Stichwörter der Autor:innen

LAFOV PET/CT[18F]FDGlow-dose [18F]FDG PETtotal-body PET/CT scanner

Autor:innen und Institutionen

  • Department of Nuclear Medicine and Clinical Molecular Imaging, University Hospital Tuebingen, 72076 Tuebingen, Germany.
  • Werner Siemens Imaging Center, Department of Preclinical Imaging and Radiopharmacy, Eberhard-Karls University Tuebingen, 72076 Tuebingen, Germany.
  • Department of Diagnostic and Interventional Radiology, University Hospital Tuebingen, 72076 Tuebingen, Germany.
  • Cluster of Excellence iFIT (EXC 2180) "Image Guided and Functionally Instructed Tumor Therapies", University of Tuebingen, 72074 Tuebingen, Germany.
  • German Cancer Consortium (DKTK), Partner Site Tuebingen, 72074 Tuebingen, Germany.

ORCID-Profile: Eduardo Calderón, Wenhong Lan, Salvador Castaneda-Vega, Andreas S Brendlin, Nils F Trautwein, Helmut Dittmann, Christian la Fougère, Lena Sophie Kiefer

Förderung

  • Deutsche Forschungsgemeinschaft
  • Faculty of Medicine, University of Tuebingen
  • Faculty of Medicine, Tuebingen, Germany

Zitieren

APA

Calderón, E., Schmidt, F. P., Lan, W., Castaneda-Vega, S., Brendlin, A. S., Trautwein, N. F., Dittmann, H., la Fougère, C., & Kiefer, L. S. (2023). Image Quality and Quantitative PET Parameters of Low-Dose [18F]FDG PET in a Long Axial Field-of-View PET/CT Scanner. Diagnostics, 13(20), 3240. https://doi.org/10.3390/diagnostics13203240

Vancouver

Calderón E, Schmidt FP, Lan W, Castaneda-Vega S, Brendlin AS, Trautwein NF, et al. Image Quality and Quantitative PET Parameters of Low-Dose [18F]FDG PET in a Long Axial Field-of-View PET/CT Scanner. Diagnostics. 2023;13(20):3240. doi:10.3390/diagnostics13203240

BibTeX
@article{Caldern2023,
  author = {Calderón, Eduardo and Schmidt, Fabian P and Lan, Wenhong and Castaneda-Vega, Salvador and Brendlin, Andreas S and Trautwein, Nils F and Dittmann, Helmut and la Fougère, Christian and Kiefer, Lena Sophie},
  title = {Image Quality and Quantitative PET Parameters of Low-Dose [18F]FDG PET in a Long Axial Field-of-View PET/CT Scanner},
  journal = {Diagnostics},
  year = {2023},
  volume = {13},
  number = {20},
  pages = {3240},
  doi = {10.3390/diagnostics13203240}
}
Kostenlos starten →