Validation of the Accuracy of Automatic Measurement of Blood Volume in Culture Bottles for Blood Culture

Kyoungbo Kim iD, Sunggyun Park iD

Diagnostics · 15.08.2023 · 13 (16) · 2685

DOI: 10.3390/diagnostics13162685 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 15.08.2023. Verfasst von Kyoungbo Kim und 1 weiteren Autor:innen. Beteiligte Einrichtung: Departments of Laboratory Medicine, Keimyung University School of Medicine, Daegu 42601, Republic of Korea. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Several manufacturers have developed systems that automatically measure the amount of blood in culture bottles. We compared the volumes measured automatically by the Virtuo instrument (bioMérieux, France) (height-based volumes) and those calculated by weighing the bottles. In all, 150 pairs of aerobic and anaerobic blood culture bottles (BacT/ALERT FA/FN Plus, bioMérieux) were randomly selected over two periods to compare the height- and weight-based volumes and analyze the effect of foam. We also estimated the limit of detection (LOD) and the cut-off value for 5 mL equine blood. The mean height-based volume was approximately 0.67 mL greater than the weight-based volume, particularly in anaerobic culture bottles. Foam did not have a significant effect. The LOD for the automatic height-based volume of equine blood was 0.2–0.4 mL. The 5 mL cut-off was 4–4.2 mL. Therefore, when reporting or monitoring blood volume within culture bottles in the laboratory, these performance characteristics should be adequately considered.

Schlagwörter

Stichwörter der Autor:innen

blood cultureblood volumebloodstream infection

Autor:innen und Institutionen

  • Departments of Laboratory Medicine, Keimyung University School of Medicine, Daegu 42601, Republic of Korea.

ORCID-Profile: Kyoungbo Kim, Sunggyun Park

Zitieren

APA

Kim, K., & Park, S. (2023). Validation of the Accuracy of Automatic Measurement of Blood Volume in Culture Bottles for Blood Culture. Diagnostics, 13(16), 2685. https://doi.org/10.3390/diagnostics13162685

Vancouver

Kim K, Park S. Validation of the Accuracy of Automatic Measurement of Blood Volume in Culture Bottles for Blood Culture. Diagnostics. 2023;13(16):2685. doi:10.3390/diagnostics13162685

BibTeX
@article{Kim2023,
  author = {Kim, Kyoungbo and Park, Sunggyun},
  title = {Validation of the Accuracy of Automatic Measurement of Blood Volume in Culture Bottles for Blood Culture},
  journal = {Diagnostics},
  year = {2023},
  volume = {13},
  number = {16},
  pages = {2685},
  doi = {10.3390/diagnostics13162685}
}
Kostenlos starten →