Preanalytical Errors in a Hematology Laboratory: An Experience from a Tertiary Care Center
Diagnostics · 06.02.2023 · 13 (4) · 591
DOI: 10.3390/diagnostics13040591 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 06.02.2023. Verfasst von Mohammad Shahid Iqbal und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, College of Applied Medical Sciences, Umm al Qura University, Makkah 21955, Saudi Arabia. Die Arbeit wurde bisher 39-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Laboratory errors arise at any stage of testing. Detecting these inaccuracies before results are revealed might delay diagnosis and treatment, causing patient distress. Here, we studied the preanalytical errors in a hematology laboratory.
Methods
This one-year retrospective analysis was conducted at the laboratory of a tertiary care hospital and included information on blood samples that were taken for hematology tests from both outpatients and inpatients. Laboratory records included sample collection and rejection information. The type and frequency of preanalytical errors were expressed as a proportion of total errors and sample number. Microsoft Excel was utilized to enter data. The results were presented in the form of frequency tables.
Results
This research included 67,892 hematology samples. For preanalytical errors, 886 samples (1.3%) were discarded. The most common preanalytical error was insufficient sample (54.17%), and the least common was an empty/damaged tube (0.4%). Erroneous samples in the emergency department were mostly insufficient and clotted, whereas pediatric sample errors were caused by insufficient and diluted samples.
Conclusion
Inadequate samples and clotted samples account for the vast majority of preanalytical factors. Insufficiency and dilutional errors were most frequent from pediatric patients. Adherence to best laboratory practices can drastically cut down on preanalytical errors.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Laboratory Medicine, College of Applied Medical Sciences, Umm al Qura University, Makkah 21955, Saudi Arabia.
- Department of Laboratory Medicine, Faculty of Applied Medical Sciences, Umm al Qura University, Makkah 21955, Saudi Arabia.
ORCID-Profile: Mohammad Shahid Iqbal, Ahmed H Qasem, Adel G Elshemi, Hibah Almasmoum
Zitieren
APA
Iqbal, M. S., Tabassum, A., Arbaeen, A. F., Qasem, A. H., Elshemi, A. G., & Almasmoum, H. (2023). Preanalytical Errors in a Hematology Laboratory: An Experience from a Tertiary Care Center. Diagnostics, 13(4), 591. https://doi.org/10.3390/diagnostics13040591
Vancouver
Iqbal MS, Tabassum A, Arbaeen AF, Qasem AH, Elshemi AG, Almasmoum H. Preanalytical Errors in a Hematology Laboratory: An Experience from a Tertiary Care Center. Diagnostics. 2023;13(4):591. doi:10.3390/diagnostics13040591
BibTeX
@article{Iqbal2023,
author = {Iqbal, Mohammad Shahid and Tabassum, Aisha and Arbaeen, Ahmad Fawzi and Qasem, Ahmed H and Elshemi, Adel G and Almasmoum, Hibah},
title = {Preanalytical Errors in a Hematology Laboratory: An Experience from a Tertiary Care Center},
journal = {Diagnostics},
year = {2023},
volume = {13},
number = {4},
pages = {591},
doi = {10.3390/diagnostics13040591}
}
EUVIMED