Phantom Quantification of Magnetoencephalography Source Imaging Distortion Caused by Deep Brain Stimulation
Brain Sciences · 22.05.2026 · 16 (6) · 554
DOI: 10.3390/brainsci16060554 ↗Kurz gefasst
Fachartikel in der Zeitschrift Brain Sciences, online erschienen am 22.05.2026. Verfasst von Saar Kariv und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurological Surgery, The University of Texas Southwestern Medical Center, Dallas, TX 75390, USA. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: the National Institute of Neurological Disease and Stroke of the National Institutes of Health.
Abstract (Originalsprache)
Objective
Previous studies of deep brain stimulation-related (DBS) artifacts in magnetoencephalography (MEG) have largely focused on the sensor level. In contrast, far less is known about their effects at the source level, where neuroscientific interpretations are typically derived. This study aims to quantify how DBS artifacts distort source-level MEG imaging.
Methods
The study used a phantom-based experimental setup to assess dipole-fitting accuracy while systematically varying the stimulation amplitude, DBS electrode configuration, and the distance between the dipole and the DBS electrode.
Results
Dipole location, angle, and amplitude errors remained within modest ranges, with the largest location and angle errors occurring at 5 mA ring-electrode stimulation (6.19 mm and 8.31 deg, respectively) and the largest amplitude errors at 15 mA ring electrodes (13.05 nAm). Location and angle errors increased significantly as the dipole moved closer to the DBS electrode, while amplitude error showed no such relationship. Continuous head position indicator coil signal quality remained stable and reliable at DBS on condition, compared to DBS off.
Conclusions
The stimulation itself does not significantly impair MEG dipole estimation, as fitting errors are similar with DBS on and off. The study introduces a quantitative framework to systematically assess DBS-related distortion via dipole-fitting error, which can also be extended to evaluate noise from other implanted or external devices.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Neurological Surgery, The University of Texas Southwestern Medical Center, Dallas, TX 75390, USA.
- Department of Radiology, The University of Texas Southwestern Medical Center, Dallas, TX 75390, USA.
- Magnetoencephalography Center of Excellence, University of Texas Southwestern Medical Center, Dallas, TX 75390, USA.
- Department of Neurology, The University of Texas Southwestern Medical Center, Dallas, TX 75390, USA.
ORCID-Profile: Jeong Woo Choi, Amy L Proskovec, Mahak Virlley, Nader Pouratian, Elizabeth M Davenport
Förderung
- the National Institute of Neurological Disease and Stroke of the National Institutes of Health
Zitieren
APA
Kariv, S., Choi, J. W., Proskovec, A. L., Virlley, M., Pruitt, T., Pouratian, N., & Davenport, E. M. (2026). Phantom Quantification of Magnetoencephalography Source Imaging Distortion Caused by Deep Brain Stimulation. Brain Sciences, 16(6), 554. https://doi.org/10.3390/brainsci16060554
Vancouver
Kariv S, Choi JW, Proskovec AL, Virlley M, Pruitt T, Pouratian N, et al. Phantom Quantification of Magnetoencephalography Source Imaging Distortion Caused by Deep Brain Stimulation. Brain Sciences. 2026;16(6):554. doi:10.3390/brainsci16060554
BibTeX
@article{Kariv2026,
author = {Kariv, Saar and Choi, Jeong Woo and Proskovec, Amy L and Virlley, Mahak and Pruitt, Tyrell and Pouratian, Nader and Davenport, Elizabeth M},
title = {Phantom Quantification of Magnetoencephalography Source Imaging Distortion Caused by Deep Brain Stimulation},
journal = {Brain Sciences},
year = {2026},
volume = {16},
number = {6},
pages = {554},
doi = {10.3390/brainsci16060554}
}
EUVIMED