Operational Flaws in Field-Effect Transistor (FET)-Based Biosensors: Gate-Leakage Current in Water-Gate FET Devices and Fake Field-Effect Signals from Biomolecular Charge
Biosensors · 23.11.2026 · 16 (12) · 681
DOI: 10.3390/bios16120681 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors, online erschienen am 23.11.2026. Verfasst von Kook-Nyung Lee. Die Arbeit nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministry of SMEs and Startups, Ministry of Trade, Industry and Energy.
Abstract (Originalsprache)
Since the possibility of direct, label-free detection of biomolecules using field-effect transistors (FETs) was first proposed around 2000, hundreds to tens of thousands of studies on nanowire/nanoribbon FET (nanoFET) biosensors have followed; yet, aside from the ion-sensitive FET (ISFET) used for pH sensing and FET-based ion sensors employing ion-selective membranes, no FET biosensor has reached actual clinical or commercial use over the past two decades. Building on prior equivalent-circuit and “fake response” analyses of nanoFET read-out, this study re-examines the electrical measurement configuration of nanoFET biosensors and shows that a substantial portion of previously reported “direct detection” signals can be explained not by the target’s own electric field but by gate-leakage current inherent to water-gate architectures and by electrical-double-layer (EDL) potential changes that accompany routine solution exchange. Without a defect-free gate dielectric, the source-drain current necessarily contains a gate-leakage component indistinguishable from the channel-resistance signal; even with an intact gate oxide, the signal is governed not by the charge of the adsorbed target molecules themselves but by the EDL interfacial potential, which is altered by the electrolyte’s ionic strength, pH, and surface chemistry. In nanoFET biosensor research, given that a buffer exchange cannot be avoided in the experimental procedure, the EDL potential change attributable to the target molecule’s own contribution cannot be quantified or converted into target-concentration information—and this difficulty is especially acute near LOD-level concentrations, where it dilutes the very advantage the sensing principle is meant to offer. It is important to understand both the difficulty of distinguishing this from a fake signal arising from gate leakage and why the FET channel-conductance shift itself is difficult to convert into a measure of the target concentration’s contribution alone, given the procedural steps the detection process requires. This is because the target-detection experimental procedure for FET biosensors inherently cannot avoid solution exchange, so the change in EDL potential distribution attributable to target-molecule adsorption cannot be isolated.
Förderung
- Ministry of SMEs and Startups
- Ministry of Trade, Industry and Energy
Zitieren
APA
Lee, K. (2026). Operational Flaws in Field-Effect Transistor (FET)-Based Biosensors: Gate-Leakage Current in Water-Gate FET Devices and Fake Field-Effect Signals from Biomolecular Charge. Biosensors, 16(12), 681. https://doi.org/10.3390/bios16120681
Vancouver
Lee K. Operational Flaws in Field-Effect Transistor (FET)-Based Biosensors: Gate-Leakage Current in Water-Gate FET Devices and Fake Field-Effect Signals from Biomolecular Charge. Biosensors. 2026;16(12):681. doi:10.3390/bios16120681
BibTeX
@article{Lee2026,
author = {Lee, Kook-Nyung},
title = {Operational Flaws in Field-Effect Transistor (FET)-Based Biosensors: Gate-Leakage Current in Water-Gate FET Devices and Fake Field-Effect Signals from Biomolecular Charge},
journal = {Biosensors},
year = {2026},
volume = {16},
number = {12},
pages = {681},
doi = {10.3390/bios16120681}
}
EUVIMED