EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Continuous Cuffless Blood Pressure Estimation Using a Multimodal Conv1D-LSTM Architecture and the SOUNDI® Wearable Device

Dario Bovio, Alice Turrini, Alice Prandoni, Alberto Porro, Pietro Cerveri, Caterina Salito iD

Biosensors · 30.07.2026 · 16 (8) · 412

DOI: 10.3390/bios16080412 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors, online erschienen am 30.07.2026. Verfasst von Dario Bovio und 5 weiteren Autor:innen. Beteiligte Einrichtung: Biocubica Srl, 20154 Milan, Italy. Hauptthemen laut MEDLINE: Wearable Electronic Devices, Blood Pressure, Blood Pressure Determination. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Arterial blood pressure (BP) is a crucial physiological parameter reflecting cardiovascular function and exhibiting dynamic oscillations in response to physiological, psychological, and environmental changes. Since conventional cuff-based measurements provide only intermittent “snapshot” readings, there is growing interest in reliable cuffless and noninvasive systems capable of capturing BP dynamics over time. This work proposes a continuous cuffless BP estimation system based on SOUNDI®, a proprietary wearable device for multimodal physiological and environmental monitoring developed by Biocubica srl. An end-to-end neural network architecture combining convolutional and long short-term memory (LSTM) layers was designed to automatically extract features from pre-processed signals and directly predict systolic and diastolic BP (SBP and DBP). Data were collected from 20 healthy subjects (7 females, 13 males) using a tailored acquisition protocol with reference targets acquired through the validated GIMA® ABPM50 device. Model performance was evaluated using both a Leave-One-Subject-Out (LOSO) cross-validation framework to assess subject-independent generalizability and a conventional 80/20 segment-level split reflecting a personalized monitoring setup. Under the subject-independent LOSO cross-validation, the model achieved a best-subject MAE of 4.93 mmHg for SBP and 3.76 mmHg for DBP, demonstrating the feasibility of generalizing across unseen individuals despite inter-subject physiological variability. Under the conventional 80/20 split, the framework achieved an MAE of 2.98 mmHg for SBP and 2.68 mmHg for DBP, with RMSE values of 4.16 mmHg and 3.72 mmHg, respectively, and Pearson correlation coefficients of R = 0.95 for SBP and R = 0.94 for DBP. These results establish a promising feasibility proof-of-concept for multimodal cuffless BP tracking under dynamic states and nocturnal conditions, highlighting both the strength of multi-sensor fusion and the value of lightweight calibration for real-world deployment on larger, clinical cohorts.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Wearable Electronic DevicesBlood PressureBlood Pressure DeterminationHumansFemaleLong Short Term MemoryMaleNeural Networks, ComputerAdult

Stichwörter der Autor:innen

LSTM neural networkblood pressure monitoringcardiovascularcufflessdeep convolutional networksmultivariate signalswearable

Autor:innen und Institutionen

  • Biocubica Srl, 20154 Milan, Italy.
  • Department of Electronics, Information and Bioengineering, Politecnico di Milano, 20133 Milan, Italy.
  • Department of Industrial, and Information Engineering, University of Pavia, 27100 Pavia, Italy.

ORCID-Profile: Caterina Salito

Zitieren

APA

Bovio, D., Turrini, A., Prandoni, A., Porro, A., Cerveri, P., & Salito, C. (2026). Continuous Cuffless Blood Pressure Estimation Using a Multimodal Conv1D-LSTM Architecture and the SOUNDI® Wearable Device. Biosensors, 16(8), 412. https://doi.org/10.3390/bios16080412

Vancouver

Bovio D, Turrini A, Prandoni A, Porro A, Cerveri P, Salito C. Continuous Cuffless Blood Pressure Estimation Using a Multimodal Conv1D-LSTM Architecture and the SOUNDI® Wearable Device. Biosensors. 2026;16(8):412. doi:10.3390/bios16080412

BibTeX
@article{Bovio2026,
  author = {Bovio, Dario and Turrini, Alice and Prandoni, Alice and Porro, Alberto and Cerveri, Pietro and Salito, Caterina},
  title = {Continuous Cuffless Blood Pressure Estimation Using a Multimodal Conv1D-LSTM Architecture and the SOUNDI® Wearable Device},
  journal = {Biosensors},
  year = {2026},
  volume = {16},
  number = {8},
  pages = {412},
  doi = {10.3390/bios16080412}
}
Kostenlos starten →