EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Advancing Lateral Flow Detection in CRISPR/Cas12a Systems Through Rational Understanding and Design Strategies of Reporter Interactions

Irina V Safenkova iD, Maria V Kamionskaya iD, Dmitriy V Sotnikov iD, Sergey F Biketov iD, Anatoly V Zherdev iD, Boris B Dzantiev iD

Biosensors · 2025

DOI: 10.3390/bios15120812 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors, 2025. Verfasst von Irina V Safenkova und 5 weiteren Autor:innen. Beteiligte Einrichtung: A.N. Bach Institute of Biochemistry, Research Centre of Biotechnology of the Russian Academy of Sciences, 119071 Moscow, Russia. Hauptthemen laut MEDLINE: CRISPR-Cas Systems, Biosensing Techniques. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Russian Science Foundation, State assignment of the State Research Center of Applied Microbiology and Biotechnology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

CRISPR/Cas12a systems coupled with lateral flow tests (LFTs) are a promising route to rapid, instrument-free nucleic acid diagnostics due to conversion target recognition into a simple visual readout via cleavage of dual-labeled single-stranded DNA reporters. However, the conventional CRISPR/Cas12a–LFT system is constructed in a format where the intact reporter should block nanoparticle conjugate migration and can produce false-positive signals and shows strong dependence on component stoichiometry and kinetics. Here, we present the first combined experimental and theoretical analysis quantifying these limitations and defining practical solutions. The experimental evaluation included 480 variants of LFT configuration with reporters differing in the concentration of interacting components and the kinetic conditions of the interactions. The most influential factor leading to 100% false-positive results was insufficient interaction time between the components; pre-incubation of the conjugate with the reporter for 5 min eliminated these artifacts. Theoretical analysis of the LFT kinetics based on a mathematical model confirmed kinetic constraints at interaction times below a few minutes, which affect the detectable signal. Reporter concentration and conjugate architecture represented the second major factors: lowering reporter concentration to 20 nM and using smaller gold nanoparticles with multivalent fluorescent reporters markedly improved sensitivity. The difference in sensitivity between various LFT configurations exceeded 50-fold. The combination of identified strategies eliminated false-positive reactions and enabled the detection of up to 20 pM of DNA target (the hisZ gene of Erwinia amylovora, a bacterial phytopathogen). The strategies reported here are general and readily transferable to other DNA targets and CRISPR/Cas12a amplification-free diagnostics.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

CRISPR-Cas SystemsBiosensing TechniquesMetal NanoparticlesGoldKinetics

Stichwörter der Autor:innen

CRISPR/Cas12aErwinia amylovoralateral flow strippoint-of-care testingreportertrans-target

Substanzen

Gold

Autor:innen und Institutionen

  • A.N. Bach Institute of Biochemistry, Research Centre of Biotechnology of the Russian Academy of Sciences, 119071 Moscow, Russia.
  • State Research Center for Applied Microbiology & Biotechnology, 142279 Obolensk, Russia.

ORCID-Profile: Irina V Safenkova, Maria V Kamionskaya, Dmitriy V Sotnikov, Sergey F Biketov, Anatoly V Zherdev, Boris B Dzantiev

Förderung

  • Russian Science Foundation
  • State assignment of the State Research Center of Applied Microbiology and Biotechnology

Zitieren

APA

Safenkova, I. V., Kamionskaya, M. V., Sotnikov, D. V., Biketov, S. F., Zherdev, A. V., & Dzantiev, B. B. (2025). Advancing Lateral Flow Detection in CRISPR/Cas12a Systems Through Rational Understanding and Design Strategies of Reporter Interactions. Biosensors. https://doi.org/10.3390/bios15120812

Vancouver

Safenkova IV, Kamionskaya MV, Sotnikov DV, Biketov SF, Zherdev AV, Dzantiev BB. Advancing Lateral Flow Detection in CRISPR/Cas12a Systems Through Rational Understanding and Design Strategies of Reporter Interactions. Biosensors. 2025. doi:10.3390/bios15120812

BibTeX
@article{Safenkova2025,
  author = {Safenkova, Irina V and Kamionskaya, Maria V and Sotnikov, Dmitriy V and Biketov, Sergey F and Zherdev, Anatoly V and Dzantiev, Boris B},
  title = {Advancing Lateral Flow Detection in CRISPR/Cas12a Systems Through Rational Understanding and Design Strategies of Reporter Interactions},
  journal = {Biosensors},
  year = {2025},
  doi = {10.3390/bios15120812}
}
Kostenlos starten →