Accuracy Enhancement in Refractive Index Sensing via Full-Spectrum Machine Learning Modeling
Biosensors · 2025
DOI: 10.3390/bios15090582 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors, 2025. Verfasst von Majid Aalizadeh und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI 48109, USA. Hauptthemen laut MEDLINE: Machine Learning, Biosensing Techniques, Refractometry. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 52 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: U.S. National Science Foundation, National Science Foundation.
Abstract (Originalsprache)
We present a full-spectrum machine learning framework for refractive index sensing using simulated absorption spectra from meta-grating structures composed of titanium or silicon nanorods under TE and TM polarizations. Linear regression was applied to 80 principal components extracted from each spectrum, and model performance was assessed using five-fold cross-validation, simulating real-world biosensing scenarios where unknown patient samples are predicted based on standard calibration data. Titanium-based structures, dominated by broadband intensity changes, yielded the lowest mean squared errors and the highest accuracy improvements—up to an 8128-fold reduction compared to the best single-feature model. In contrast, silicon-based structures, governed by narrow resonances, showed more modest gains due to spectral nonlinearity that limits the effectiveness of global linear models. We also show that even the best single-wavelength predictor is identified through data-driven analysis, not visual selection, highlighting the value of automated feature preselection. These findings demonstrate that spectral shape plays a key role in modeling performance and that full-spectrum linear approaches are especially effective for intensity-modulated index sensors.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI 48109, USA.
- Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor, MI 48109, USA.
- Center for Wireless Integrated MicroSensing and Systems (WIMS2), University of Michigan, Ann Arbor, MI 48109, USA.
- Max Harry Weil Institute for Critical Care Research and Innovation, University of Michigan, Ann Arbor, MI 48109, USA.
- Department of Computational Medicine and Bioinformatics, University of Michigan, Ann Arbor, MI 48109, USA.
- Informatics Institute, Istanbul Technical University, Istanbul 34485, Turkey.
ORCID-Profile: Majid Aalizadeh, Chinmay Raut, Xudong Fan
Förderung
- U.S. National Science Foundation
- National Science Foundation
Zitieren
APA
Aalizadeh, M., Raut, C., Azmoudeh Afshar, M., Tabartehfarahani, A., & Fan, X. (2025). Accuracy Enhancement in Refractive Index Sensing via Full-Spectrum Machine Learning Modeling. Biosensors. https://doi.org/10.3390/bios15090582
Vancouver
Aalizadeh M, Raut C, Azmoudeh Afshar M, Tabartehfarahani A, Fan X. Accuracy Enhancement in Refractive Index Sensing via Full-Spectrum Machine Learning Modeling. Biosensors. 2025. doi:10.3390/bios15090582
BibTeX
@article{Aalizadeh2025,
author = {Aalizadeh, Majid and Raut, Chinmay and Azmoudeh Afshar, Morteza and Tabartehfarahani, Ali and Fan, Xudong},
title = {Accuracy Enhancement in Refractive Index Sensing via Full-Spectrum Machine Learning Modeling},
journal = {Biosensors},
year = {2025},
doi = {10.3390/bios15090582}
}
EUVIMED