EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

FRET-Based TURN-ON Aptasensor for the Sensitive Detection of CK-MB

Rabia Asghar, Madiha Rasheed iD, Xuefei Lv, Yulin Deng iD

Biosensors · 2025

DOI: 10.3390/bios15070446 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors, 2025. Verfasst von Rabia Asghar und 3 weiteren Autor:innen. Beteiligte Einrichtung: Beijing Key Laboratory for Separation and Analysis in Biomedicine and Pharmaceuticals, School of Medical Technology, Beijing Institute of Technology,…. Hauptthemen laut MEDLINE: Fluorescence Resonance Energy Transfer, Biosensing Techniques, Aptamers, Nucleotide. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Beijing Key Laboratory for Separation and Analysis in Biomedicine and Pharmaceuticals.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A fluorescent sandwich assay was devised to quantify CK-MB. In a typical immunoassay, antibodies bind to the target, and the detected signal is quantified according to the target’s concentration. We innovated a unique fluorescence assay known as the “enzyme-linked aptamer assay” (ELAA) by substituting antibodies with a pair of high-affinity aptamers labelled with biotin, namely apt. A1 and apt. A2. Avidin-labelled ALP binds to biotin-labelled aptamers, hydrolyzing its substrate, 2-phosphoascorbic acid trisodium salt, resulting in the formation of ascorbic acid. The catalytic hydrolysate functions as a reducing agent, causing the deterioration of MoS2 nanosheets. This results in the transformation of MoS2 nanosheets into nanoribbons, leading to the release of quenched AGQDs. The reestablishment of fluorescence is triggered by Förster Resonance Energy Transfer (FRET) between the MoS2 nanoribbons and AGQDs, enhancing the sensitivity of disease biomarker detection. The working range for detection falls between 2.5 nM and 160 nM, and the limit of detection (LOD) for CK-MB is verified at 0.20 nM.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Fluorescence Resonance Energy TransferBiosensing TechniquesAptamers, NucleotideHumansLimit of DetectionMolybdenumImmunoassayDisulfides

Stichwörter der Autor:innen

AGQDs@MoS2 nanoassemblyALPFRETMoS2 nanosheetsaminated graphene quantum dots (AGQDs)aptameraptasensor

Substanzen

Aptamers, NucleotideMolybdenummolybdenum disulfideDisulfides

Autor:innen und Institutionen

  • Beijing Key Laboratory for Separation and Analysis in Biomedicine and Pharmaceuticals, School of Medical Technology, Beijing Institute of Technology, Beijing 100081, China.
  • BIT & GS Technology Co., Ltd., Beijing 100043, China.

ORCID-Profile: Madiha Rasheed, Yulin Deng

Förderung

  • Beijing Key Laboratory for Separation and Analysis in Biomedicine and Pharmaceuticals

Zitieren

APA

Asghar, R., Rasheed, M., Lv, X., & Deng, Y. (2025). FRET-Based TURN-ON Aptasensor for the Sensitive Detection of CK-MB. Biosensors. https://doi.org/10.3390/bios15070446

Vancouver

Asghar R, Rasheed M, Lv X, Deng Y. FRET-Based TURN-ON Aptasensor for the Sensitive Detection of CK-MB. Biosensors. 2025. doi:10.3390/bios15070446

BibTeX
@article{Asghar2025,
  author = {Asghar, Rabia and Rasheed, Madiha and Lv, Xuefei and Deng, Yulin},
  title = {FRET-Based TURN-ON Aptasensor for the Sensitive Detection of CK-MB},
  journal = {Biosensors},
  year = {2025},
  doi = {10.3390/bios15070446}
}
Kostenlos starten →