Achievement of 15-Minute Adaptive PCR Benchmark with 1370 nm Laser Heating
Biosensors · 2025
DOI: 10.3390/bios15040258 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors, 2025. Verfasst von Nicholas Spurlock und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Vanderbilt University, Nashville, TN 37235, USA. Hauptthemen laut MEDLINE: Polymerase Chain Reaction, Lasers, Biosensing Techniques. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIAID NIH HHS, NIH HHS, East Tennessee Foundation.
Abstract (Originalsprache)
In low-resource and point-of-care settings, traditional PCR often faces challenges of poor sample preparation, adverse environmental conditions, and long assay times. We have previously described a laboratory-based instrument to achieve “adaptive” PCR, a PCR thermocycling control system that replaces preset cycling times and temperatures with the optical monitoring of added L-DNA stereoisomers matching the sequences of the reaction primers and target. These L-DNA biosensors directly monitor DNA hybridization, compensating for ambient environmental conditions and poor sample preparation. This report describes instrument simplifications and a comparative evaluation of both direct photothermal and plasmonic laser heating to reduce the assay time to 15 min. Instrument performance was assessed using a split sample design to compare reaction performances of 1370 and 808 nm adaptive PCR heating modalities to a standard PCR instrument. Both the novel 1370 nm direct heating and the 808 nm plasmonic method achieved target amplification similar to the traditional PCR system within 15 min. However, a major disadvantage of 808 nm heating was nanorod optical interference that reduced the fluorescence signal from PCR probes and optical cycling components. Further characterization of the 1370 nm direct heating method found comparable limits of detection of 100 copies/µL and reaction efficiencies of approximately 2 for both the 1370 nm system and the traditional PCR instrument. These results suggest that a field-deployable PCR instrument design incorporating both adaptive optical control and 1370 nm laser heating can achieve 15 min sample assay times without sacrificing analytical sensitivity.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Biomedical Engineering, Vanderbilt University, Nashville, TN 37235, USA.
- Department of Physics and Astronomy, Vanderbilt University, Nashville, TN 37235, USA.
ORCID-Profile: Nicholas Spurlock, William E Gabella, Kunal Chugh, Frederick R Haselton
Förderung
- NIAID NIH HHS
- NIH HHS
- East Tennessee Foundation
- NIH
- East Tennessee Foundation Cobb Scholarship
Zitieren
APA
Spurlock, N., Alfaro, R., Gabella, W. E., Chugh, K., Pask, M. E., Baudenbacher, F., & Haselton, F. R. (2025). Achievement of 15-Minute Adaptive PCR Benchmark with 1370 nm Laser Heating. Biosensors. https://doi.org/10.3390/bios15040258
Vancouver
Spurlock N, Alfaro R, Gabella WE, Chugh K, Pask ME, Baudenbacher F, et al. Achievement of 15-Minute Adaptive PCR Benchmark with 1370 nm Laser Heating. Biosensors. 2025. doi:10.3390/bios15040258
BibTeX
@article{Spurlock2025,
author = {Spurlock, Nicholas and Alfaro, Rosana and Gabella, William E and Chugh, Kunal and Pask, Megan E and Baudenbacher, Franz and Haselton, Frederick R},
title = {Achievement of 15-Minute Adaptive PCR Benchmark with 1370 nm Laser Heating},
journal = {Biosensors},
year = {2025},
doi = {10.3390/bios15040258}
}
EUVIMED