Analysis and Tracking of Intra-Needle Ultrasound Pleural Signals for Improved Anesthetic Procedures in the Thoracic Region

Fu-Wei Su, Chia-Wei Yang, Ching-Fang Yang, Yi-En Tsai, Wei-Nung Teng, Huihua Kenny Chiang iD

Biosensors · 21.03.2025 · 15 (4) · 201

DOI: 10.3390/bios15040201 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors, online erschienen am 21.03.2025. Verfasst von Fu-Wei Su und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology, Taipei Veterans General Hospital, Taipei City 11217, Taiwan. Hauptthemen laut MEDLINE: Pleura, Thorax. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministry of Science and Technology, Taiwan, National Science and Technology Council, Taiwan.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Ultrasonography is commonly employed during thoracic regional anesthesia; however, its accuracy can be affected by factors such as obesity and poor penetration through the rib window. Needle-sized ultrasound transducers, known as intra-needle ultrasound (INUS) transducers, have been developed to detect the pleura and fascia using a one-dimensional radio frequency mode ultrasound signal. In this study, we aimed to use time-frequency analysis to characterize the pleural signal and develop an automated tool to identify the pleura during medical procedures.

Methods

We developed an INUS system and investigated the pleural signal it measured by establishing a phantom study, and an in vivo animal study. Signals from the pleura, endothoracic fascia, and intercostal muscles were analyzed. Additionally, we conducted time- and frequency-domain analyses of the pleural and alveolar signals.

Results

We identified the unique characteristics of the pleura, including a flickering phenomenon, speckle-like patterns, and highly variable multi-band spectra in the ultrasound signal during the breathing cycle. These characteristics are likely due to the multiple reflections from the sliding visceral pleura and alveoli. This automated identification of the pleura can enhance the safety for thoracic regional anesthesia, particularly in difficult cases.

Conclusions

The unique flickering pleural signal based on INUS can be processed by time-frequency domain analysis and further tracked by an auto-identification algorithm. This technique has potential applications in thoracic regional anesthesia and other interventions. However, further studies are required to validate this hypothesis. Key Points Summary: Question: How can the ultrasound pleural signal be distinguished from other tissues during breathing?

Findings

The frequency domain analysis of the pleural ultrasound signal showed fast variant and multi-band characteristics. We suggest this is due to ultrasound distortion caused by the interface of multiple moving alveoli. The multiple ultrasonic reflections from the sliding pleura and alveoli returned in variable and multi-banded frequency. Meaning: The distinguished pleural signal can be used for the auto-identification of the pleura for further clinical respiration monitoring and safety during regional anesthesia. Glossary of Terms: intra-needle ultrasound (INUS); radio frequency (RF); short-time Fourier transform (STFT); intercostal nerve block (ICNB); paravertebral block (PVB); pulse repetition frequency (PRF).

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PleuraThoraxAnimalsUltrasonographyNeedlesHumansPhantoms, ImagingSwine

Stichwörter der Autor:innen

pleurasignal analysisultrasound

Autor:innen und Institutionen

  • Department of Anesthesiology, Taipei Veterans General Hospital, Taipei City 11217, Taiwan.
  • Department of Biomedical Engineering, National Yang-Ming Chiao-Tung University, Taipei City 112304, Taiwan.
  • School of Medicine, National Yang-Ming Chiao-Tung University, Taipei City 112304, Taiwan.

ORCID-Profile: Huihua Kenny Chiang

Förderung

  • Ministry of Science and Technology, Taiwan
  • National Science and Technology Council, Taiwan

Zitieren

APA

Su, F. W., Yang, C. W., Yang, C. F., Tsai, Y. E., Teng, W. N., & Chiang, H. K. (2025). Analysis and Tracking of Intra-Needle Ultrasound Pleural Signals for Improved Anesthetic Procedures in the Thoracic Region. Biosensors, 15(4), 201. https://doi.org/10.3390/bios15040201

Vancouver

Su FW, Yang CW, Yang CF, Tsai YE, Teng WN, Chiang HK. Analysis and Tracking of Intra-Needle Ultrasound Pleural Signals for Improved Anesthetic Procedures in the Thoracic Region. Biosensors. 2025;15(4):201. doi:10.3390/bios15040201

BibTeX
@article{Su2025,
  author = {Su, Fu-Wei and Yang, Chia-Wei and Yang, Ching-Fang and Tsai, Yi-En and Teng, Wei-Nung and Chiang, Huihua Kenny},
  title = {Analysis and Tracking of Intra-Needle Ultrasound Pleural Signals for Improved Anesthetic Procedures in the Thoracic Region},
  journal = {Biosensors},
  year = {2025},
  volume = {15},
  number = {4},
  pages = {201},
  doi = {10.3390/bios15040201}
}
Kostenlos starten →