Low-Cost, Open-Source, High-Precision Pressure Controller for Multi-Channel Microfluidics
Biosensors · 2025
DOI: 10.3390/bios15030154 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors, 2025. Verfasst von Mart Ernits und 4 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Technology, University of Tartu, Nooruse 1, 50411 Tartu, Estonia. Hauptthemen laut MEDLINE: Microfluidics, Microfluidic Analytical Techniques, Lab-On-A-Chip Devices. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Union's Horizon 2020 research and innovation programme, JPIAMR - Joint Programming Initiative on Antimicrobial Resistance, European Union’s Horizon 2020 research and innovation programme.
Abstract (Originalsprache)
Microfluidics is a technology that manipulates liquids on the scales ranging from microliters to femtoliters. Such low volumes require precise control over pressures that drive their flow into the microfluidic chips. This article describes a custom-built pressure controller for driving microfluidic chips. The pressure controller features piezoelectrically controlled pressure regulation valves. As an open-source system, it offers high customizability and allows users to modify almost every aspect. The cost is roughly a third of what similar, alternative, commercially available piezoelectrically controlled pressure regulators could be purchased for. The measured output pressure values of the device vary less than 0.7% from the device’s reported pressure values when the requested pressure is between −380 and 380 mbar. Importantly, the output pressure the device creates fluctuates only ±0.2 mbar when the pressure is cycled between 10 and 500 mbar. The pressure reading accuracy and stability validation suggest that the device is highly feasible for many advanced (low-pressure) microfluidic applications. Here, we compare the main features of our device to commercially and non-commercially available alternatives and further demonstrate the device’s performance and accessibility in successful microfluidic hydrodynamic focusing (MHF)-based synthesis of large unilamellar vesicles (LUVs).
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Technology, University of Tartu, Nooruse 1, 50411 Tartu, Estonia.
- Department of Bioproducts and Biosystems, Aalto University School of Chemical Engineering, Kemistintie 1, 02150 Espoo, Finland.
ORCID-Profile: Olavi Reinsalu, Andreas Kyritsakis, Veikko Linko
Förderung
- European Union's Horizon 2020 research and innovation programme
- JPIAMR - Joint Programming Initiative on Antimicrobial Resistance
- European Union’s Horizon 2020 research and innovation programme
- JPIAMR—the Joint Programming Initiative on Antimicrobial Resistance
Zitieren
APA
Ernits, M., Reinsalu, O., Kyritsakis, A., Linko, V., & Zadin, V. (2025). Low-Cost, Open-Source, High-Precision Pressure Controller for Multi-Channel Microfluidics. Biosensors. https://doi.org/10.3390/bios15030154
Vancouver
Ernits M, Reinsalu O, Kyritsakis A, Linko V, Zadin V. Low-Cost, Open-Source, High-Precision Pressure Controller for Multi-Channel Microfluidics. Biosensors. 2025. doi:10.3390/bios15030154
BibTeX
@article{Ernits2025,
author = {Ernits, Mart and Reinsalu, Olavi and Kyritsakis, Andreas and Linko, Veikko and Zadin, Veronika},
title = {Low-Cost, Open-Source, High-Precision Pressure Controller for Multi-Channel Microfluidics},
journal = {Biosensors},
year = {2025},
doi = {10.3390/bios15030154}
}
EUVIMED