Combined Dielectric-Optical Characterization of Single Cells Using Dielectrophoresis-Imaging Flow Cytometry
Biosensors · 2024
DOI: 10.3390/bios14120577 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors, 2024. Verfasst von Behnam Arzhang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Computer Engineering, University of Manitoba, Winnipeg, MB R3T 5V6, Canada. Hauptthemen laut MEDLINE: Flow Cytometry, Cricetulus, Electrophoresis. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 66 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Sciences and Engineering Research Council.
Abstract (Originalsprache)
In this paper, we present a microfluidic flow cytometer for simultaneous imaging and dielectric characterization of individual biological cells within a flow. Utilizing a combination of dielectrophoresis (DEP) and high-speed imaging, this system offers a dual-modality approach to analyze both cell morphology and dielectric properties, enhancing the ability to analyze, characterize, and discriminate cells in a heterogeneous population. A high-speed camera is used to capture images of and track multiple cells in real-time as they flow through a microfluidic channel. A wide channel is used, enabling analysis of many cells in parallel. A coplanar electrode array perpendicular to cell flow is incorporated at the bottom of the channel to perform dielectrophoresis-based dielectric characterization. A frequency-dependent voltage applied to the array produces a non-uniform electric field, translating cells to higher or lower velocity depending on their dielectric polarizability. In this paper, we demonstrate how cell size, obtained by optical imaging, and DEP response, obtained by particle tracking, can be used to discriminate viable and non-viable Chinese hamster ovary cells in a heterogeneous cell culture. Multiphysics electrostatic-fluid dynamics simulation is used to develop a relationship between cell incoming velocity, differential velocity, size, and the cell’s polarizability, which can subsequently be used to evaluate its physiological state. Measurement of a mixture of polystyrene microspheres is used to evaluate the accuracy of the cytometer.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electrical and Computer Engineering, University of Manitoba, Winnipeg, MB R3T 5V6, Canada.
- National Institute for Bioprocessing Research and Training, A94 X099 Dublin, Ireland.
ORCID-Profile: Behnam Arzhang, Emerich Kovacs, Douglas J Thomson, Greg E Bridges
Förderung
- Natural Sciences and Engineering Research Council
Zitieren
APA
Arzhang, B., Lee, J., Kovacs, E., Butler, M., Salimi, E., Thomson, D. J., & Bridges, G. E. (2024). Combined Dielectric-Optical Characterization of Single Cells Using Dielectrophoresis-Imaging Flow Cytometry. Biosensors. https://doi.org/10.3390/bios14120577
Vancouver
Arzhang B, Lee J, Kovacs E, Butler M, Salimi E, Thomson DJ, et al. Combined Dielectric-Optical Characterization of Single Cells Using Dielectrophoresis-Imaging Flow Cytometry. Biosensors. 2024. doi:10.3390/bios14120577
BibTeX
@article{Arzhang2024,
author = {Arzhang, Behnam and Lee, Justyna and Kovacs, Emerich and Butler, Michael and Salimi, Elham and Thomson, Douglas J and Bridges, Greg E},
title = {Combined Dielectric-Optical Characterization of Single Cells Using Dielectrophoresis-Imaging Flow Cytometry},
journal = {Biosensors},
year = {2024},
doi = {10.3390/bios14120577}
}
EUVIMED