Fluorescence Immunoassay of Prostate-Specific Antigen Using 3D Paddle Screw-Type Devices and Their Rotating System
Biosensors · 2024
DOI: 10.3390/bios14100494 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors, 2024. Verfasst von Su Bin Han und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pharmaceutical Engineering, Soonchunhyang University, Asan 31538, Republic of Korea. Hauptthemen laut MEDLINE: Prostate-Specific Antigen, Biosensing Techniques. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation of Korea.
Abstract (Originalsprache)
In this paper, we present a sensitive and highly reproducible fluorescence immunosensor for detecting PSA in human serum. A unique feature of this study is that it uses creatively designed paddle screw-type devices and their custom-made rotating system for PSA immunoassay. The paddle screw devices were designed to maximize the surface-to-volume ratio over which the immunoassay reaction could occur to improve detection sensitivity. This paddle screw-based immunoassay offers an accessible and efficient method with a short analysis time of less than 30 min. Active rotation of the paddle screw plays a crucial role in fast and accurate analysis of PSA. Additionally, a paddle screw-based immunoassay and subsequent fluorescence detection using a custom prototype fluorescence detection system were compared to a typical well plate-based immunoassay system. Results of PSA detection in human serum showed that the detection sensitivity through the paddle screw-based analysis improved about five times compared to that with a well plate-based analysis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Pharmaceutical Engineering, Soonchunhyang University, Asan 31538, Republic of Korea.
ORCID-Profile: Soo Suk Lee
Förderung
- National Research Foundation of Korea
Zitieren
APA
Han, S. B., Kim, H. S., Jo, Y. J., & Lee, S. S. (2024). Fluorescence Immunoassay of Prostate-Specific Antigen Using 3D Paddle Screw-Type Devices and Their Rotating System. Biosensors. https://doi.org/10.3390/bios14100494
Vancouver
Han SB, Kim HS, Jo YJ, Lee SS. Fluorescence Immunoassay of Prostate-Specific Antigen Using 3D Paddle Screw-Type Devices and Their Rotating System. Biosensors. 2024. doi:10.3390/bios14100494
BibTeX
@article{Han2024,
author = {Han, Su Bin and Kim, Han Sol and Jo, Young Ju and Lee, Soo Suk},
title = {Fluorescence Immunoassay of Prostate-Specific Antigen Using 3D Paddle Screw-Type Devices and Their Rotating System},
journal = {Biosensors},
year = {2024},
doi = {10.3390/bios14100494}
}
EUVIMED