Ultrasoft Long-Lasting Reusable Hydrogel-Based Sensor Patch for Biosignal Recording

Alexandre Tessier, Shuyun Zhuo iD, Shideh Kabiri Ameri

Biosensors · 2024

DOI: 10.3390/bios14080405 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors, 2024. Verfasst von Alexandre Tessier und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Computer Engineering, Queen's University, Kingston, ON K7L 3N6, Canada. Hauptthemen laut MEDLINE: Hydrogels, Electromyography, Electrocardiography. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Sciences and Engineering Research Council of Canada.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Here, we report an ultrasoft extra long-lasting, reusable hydrogel-based sensor that enables high-quality electrophysiological recording with low-motion artifacts. The developed sensor can be used and stored in an ambient environment for months before being reused. The developed sensor is made of a self-adhesive electrical-conductivity-enhanced ultrasoft hydrogel mounted in an Ecoflex-based frame. The hydrogel’s conductivity was enhanced by incorporating polypyrrole (PPy), resulting in a conductivity of 0.25 S m−1. Young’s modulus of the sensor is only 12.9 kPa, and it is stretchable up to 190%. The sensor was successfully used for electrocardiography (ECG) and electromyography (EMG). Our results indicate that using the developed hydrogel-based sensor, the signal-to-noise ratio of recorded electrophysiological signals was improved in comparison to that when medical-grade silver/silver chloride (Ag/AgCl) wet gel electrodes were used (33.55 dB in comparison to 22.16 dB). Due to the ultra-softness, high stretchability, and self-adhesion of the developed sensor, it can conform to the skin and, therefore, shows low susceptibility to motion. In addition, the sensor shows no sign of irritation or allergic reaction, which usually occurs after long-term wearing of medical-grade Ag/AgCl wet gel electrodes on the skin. Further, the sensor is fabricated using a low-cost and scalable fabrication process.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HydrogelsElectromyographyElectrocardiographyHumansBiosensing TechniquesElectric ConductivityPolymersPyrrolesWearable Electronic DevicesElectrodesSignal-To-Noise Ratio

Stichwörter der Autor:innen

electrophysiological recordingreusable sensorsensor-skin interfacesignal-to-noise ratioultrasoft hydrogel

Substanzen

HydrogelspolypyrrolePolymersPyrroles

Autor:innen und Institutionen

  • Department of Electrical and Computer Engineering, Queen's University, Kingston, ON K7L 3N6, Canada.
  • Centre for Neuroscience Studies, Queen's University, Kingston, ON K7L 3N6, Canada.

ORCID-Profile: Shuyun Zhuo

Förderung

  • Natural Sciences and Engineering Research Council of Canada

Zitieren

APA

Tessier, A., Zhuo, S., & Kabiri Ameri, S. (2024). Ultrasoft Long-Lasting Reusable Hydrogel-Based Sensor Patch for Biosignal Recording. Biosensors. https://doi.org/10.3390/bios14080405

Vancouver

Tessier A, Zhuo S, Kabiri Ameri S. Ultrasoft Long-Lasting Reusable Hydrogel-Based Sensor Patch for Biosignal Recording. Biosensors. 2024. doi:10.3390/bios14080405

BibTeX
@article{Tessier2024,
  author = {Tessier, Alexandre and Zhuo, Shuyun and Kabiri Ameri, Shideh},
  title = {Ultrasoft Long-Lasting Reusable Hydrogel-Based Sensor Patch for Biosignal Recording},
  journal = {Biosensors},
  year = {2024},
  doi = {10.3390/bios14080405}
}