Fachartikel in der Zeitschrift Biosensors, online erschienen am 16.07.2024. Verfasst von Pradakshina Sharma und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biotechnology, School of Chemical and Life Sciences, Jamia Hamdard, Hamdard Nagar, New Delhi 110062, India. Hauptthemen laut MEDLINE: Chikungunya virus, Silver, Biosensing Techniques, Metal Nanoparticles. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Indian Council of Medical Research, ICMR, New Delhi.
Abstract (Originalsprache)
Medical devices have progressed from their initial bulky forms to smart devices. However, their rigidity hampers their seamless integration into everyday life. The fields of stretchable, textile, and flexible electronics are emerging research areas with the potential to drive significant technological progress. This research presents a laboratory-based technique to produce highly sensitive and flexible biosensors for detecting the chikungunya virus. These biosensors are based on 0D nanomaterials and demonstrate significant advancements in voltammetry. The electrochemical platform was created utilizing the stencil printing (StPE) technique. Adapting the biosensor setup involved the selection of aptamer as the biorecognition element bound with silver nanoparticles (AgNPs). This biosensor was employed in the voltammetric identification of the Chikungunya virus antigen (CHIKV-Ag) within a solution containing 0.5 mM potassium ferro/ferri cyanide, a redox pair. The biosensor was employed to evaluate CHIKV-Ag within a human serum sample. It demonstrated a linear detection span ranging from 0.1 ng/mL to 1 μg/mL, with a detection limit of 0.1 ng/mL for CHIKV-Ag. The proposed approach, due to its flexibility in production and the electrocatalytic attributes displayed by the zero-dimensional nanostructure, presents innovative opportunities for cost-effective and tailored aptamer-based bioelectronics, thereby broadening the scope of this domain.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Biotechnology, School of Chemical and Life Sciences, Jamia Hamdard, Hamdard Nagar, New Delhi 110062, India.
- Institute of Crystallography, National Research Council, 00143 Rome, Italy.
ORCID-Profile: Roberto Pilloton
Förderung
- Indian Council of Medical Research
- ICMR, New Delhi
Zitieren
APA
Sharma, P., Hasan, M. R., Naikoo, U. M., Khatoon, S., Pilloton, R., & Narang, J. (2024). Aptamer Based on Silver Nanoparticle-Modified Flexible Carbon Ink Printed Electrode for the Electrochemical Detection of Chikungunya Virus. Biosensors, 14(7), 344. https://doi.org/10.3390/bios14070344
Vancouver
Sharma P, Hasan MR, Naikoo UM, Khatoon S, Pilloton R, Narang J. Aptamer Based on Silver Nanoparticle-Modified Flexible Carbon Ink Printed Electrode for the Electrochemical Detection of Chikungunya Virus. Biosensors. 2024;14(7):344. doi:10.3390/bios14070344
BibTeX
@article{Sharma2024,
author = {Sharma, Pradakshina and Hasan, Mohd Rahil and Naikoo, Ubaid Mushtaq and Khatoon, Shaheen and Pilloton, Roberto and Narang, Jagriti},
title = {Aptamer Based on Silver Nanoparticle-Modified Flexible Carbon Ink Printed Electrode for the Electrochemical Detection of Chikungunya Virus},
journal = {Biosensors},
year = {2024},
volume = {14},
number = {7},
pages = {344},
doi = {10.3390/bios14070344}
}
EUVIMED