EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Real-Time Monitoring of a Nucleic Acid Amplification Reaction Using a Mass Sensor Based on a Quartz-Crystal Microbalance

Hideto Kumagai, Hiroyuki Furusawa iD

Biosensors · 2024

DOI: 10.3390/bios14040155 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors, 2024. Verfasst von Hideto Kumagai und 1 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of Organic Materials Science, Yamagata University, Yonezawa 992-8510, Japan. Hauptthemen laut MEDLINE: Quartz Crystal Microbalance Techniques, Nucleic Acid Amplification Techniques, Biosensing Techniques, DNA. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Japan Science and Technology Agency, Piezo Parts Co., Ltd..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Nucleic acid amplification reactions such as polymerase chain reaction (PCR), which uses a DNA polymerase to amplify individual double-stranded DNA fragments, are a useful technique for visualizing the presence of specific genomes. Although the fluorescent labeling method is mainly used with DNA amplification, other detection methods should be considered for further improvements, such as miniaturization and cost reduction, of reaction-monitoring devices. In this study, the quartz-crystal microbalance (QCM) method, which can measure nanogram-order masses, was applied for the real-time detection of DNA fragments in a solution with nucleic acids. This was combined with an isothermal nucleic acid amplification reaction based on the recombinase polymerase amplification (RPA) method, which allowed DNA amplification at a constant temperature. When the DNA amplification reaction was initiated on a QCM sensor plate with an immobilized primer DNA strand, a significant increase in mass was observed compared to when the primer DNA was not immobilized. QCM was shown to be sufficiently sensitive for the in situ detection of amplified DNA fragments. Combining a portable QCM device and RPA offers a sensitive point-of-care method for detecting nucleic acids.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Quartz Crystal Microbalance TechniquesNucleic Acid Amplification TechniquesBiosensing TechniquesDNA

Stichwörter der Autor:innen

DNA fragment detectionisothermal nucleic acid amplification reactionmass sensorpolymerase chain reactionquartz-crystal microbalancerecombinase polymerase amplification

Substanzen

DNA

Autor:innen und Institutionen

  • Graduate School of Organic Materials Science, Yamagata University, Yonezawa 992-8510, Japan.
  • Institute for the Promotion of General Graduate Education (IPGE), Yamagata University, Yonezawa 992-8510, Japan.

ORCID-Profile: Hiroyuki Furusawa

Förderung

  • Japan Science and Technology Agency
  • Piezo Parts Co., Ltd.

Zitieren

APA

Kumagai, H., & Furusawa, H. (2024). Real-Time Monitoring of a Nucleic Acid Amplification Reaction Using a Mass Sensor Based on a Quartz-Crystal Microbalance. Biosensors. https://doi.org/10.3390/bios14040155

Vancouver

Kumagai H, Furusawa H. Real-Time Monitoring of a Nucleic Acid Amplification Reaction Using a Mass Sensor Based on a Quartz-Crystal Microbalance. Biosensors. 2024. doi:10.3390/bios14040155

BibTeX
@article{Kumagai2024,
  author = {Kumagai, Hideto and Furusawa, Hiroyuki},
  title = {Real-Time Monitoring of a Nucleic Acid Amplification Reaction Using a Mass Sensor Based on a Quartz-Crystal Microbalance},
  journal = {Biosensors},
  year = {2024},
  doi = {10.3390/bios14040155}
}
Kostenlos starten →