Noninvasive Estimation of Tumor Interstitial Fluid Pressure from Subharmonic Scattering of Ultrasound Contrast Microbubbles

Yun Wang, Huimin Lu, Laixin Huang, Deyu Li, Weibao Qiu, Lingling Li, Gang Xu, Min Su iD, Jianhua Zhou iD, Fei Li iD

Biosensors · 2023

DOI: 10.3390/bios13050528 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors, 2023. Verfasst von Yun Wang und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ultrasound, Sun Yat-sen University Cancer Center, State Key Laboratory of Oncology in South China, Collaborative Innovation Center for …. Hauptthemen laut MEDLINE: Microbubbles, Neoplasms. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, Shenzhen Science and Technology Program, Shenzhen Fundamental Research Program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The noninvasive estimation of interstitial fluid pressure (IFP) using ultrasound contrast agent (UCA) microbubbles as pressure sensors will provide tumor treatments and efficacy assessments with a promising tool. This study aimed to verify the efficacy of the optimal acoustic pressure in vitro in the prediction of tumor IFPs based on UCA microbubbles’ subharmonic scattering. A customized ultrasound scanner was used to generate subharmonic signals from microbubbles’ nonlinear oscillations, and the optimal acoustic pressure was determined in vitro when the subharmonic amplitude reached the most sensitive to hydrostatic pressure changes. This optimal acoustic pressure was then applied to predict IFPs in tumor-bearing mouse models, which were further compared with the reference IFPs measured using a standard tissue fluid pressure monitor. An inverse linear relationship and good correlation (r = −0.853, p < 0.001) existed between the subharmonic amplitude and tumor IFPs at the optimal acoustic pressure of 555 kPa, and pressure sensitivity was 1.019 dB/mmHg. No statistical differences were found between the pressures measured by the standard device and those estimated via the subharmonic amplitude, as confirmed by cross-validation (mean absolute errors from 2.00 to 3.09 mmHg, p > 0.05). Our findings demonstrated that in vitro optimized acoustic parameters for UCA microbubbles’ subharmonic scattering can be applied for the noninvasive estimation of tumor IFPs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MicrobubblesNeoplasmsAnimalsMiceContrast MediaExtracellular FluidUltrasonography

Stichwörter der Autor:innen

interstitial fluid pressuresubharmonic scatteringultrasound contrast agent microbubbles

Substanzen

Contrast Media

Autor:innen und Institutionen

  • Department of Ultrasound, Sun Yat-sen University Cancer Center, State Key Laboratory of Oncology in South China, Collaborative Innovation Center for Cancer Medicine, Guangzhou 510060, China.
  • Paul C. Lauterbur Research Center for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen 518055, China.
  • Department of Critical Care Medicine, Nanjing Drum Tower Hospital, Affiliated Hospital of Nanjing University Medical School, Nanjing 210008, China.
  • Key Laboratory for Biomechanics and Mechanobiology of the Ministry of Education, Beijing Advanced Innovation Centre for Biomedical Engineering, School of Biological Science and Medical Engineering, Beihang University, B…
  • Liver Transplant Center, Organ Transplant Center, West China Hospital of Sichuan University, Chengdu 610041, China.
  • Laboratory of Liver Transplantation, Key Laboratory of Transplant Engineering and Immunology, West China Hospital of Sichuan University, Chengdu 610093, China.

ORCID-Profile: Min Su, Jianhua Zhou, Fei Li

Förderung

  • National Natural Science Foundation of China
  • Shenzhen Science and Technology Program
  • Shenzhen Fundamental Research Program
  • Shenzhen Key Laboratory of Ultrasound Imaging and Therapy
  • China Postdoctoral Science Foundation
  • Shenzhen Basic Research Program

Zitieren

APA

Wang, Y., Lu, H., Huang, L., Li, D., Qiu, W., Li, L., Xu, G., Su, M., Zhou, J., & Li, F. (2023). Noninvasive Estimation of Tumor Interstitial Fluid Pressure from Subharmonic Scattering of Ultrasound Contrast Microbubbles. Biosensors. https://doi.org/10.3390/bios13050528

Vancouver

Wang Y, Lu H, Huang L, Li D, Qiu W, Li L, et al. Noninvasive Estimation of Tumor Interstitial Fluid Pressure from Subharmonic Scattering of Ultrasound Contrast Microbubbles. Biosensors. 2023. doi:10.3390/bios13050528

BibTeX
@article{Wang2023,
  author = {Wang, Yun and Lu, Huimin and Huang, Laixin and Li, Deyu and Qiu, Weibao and Li, Lingling and Xu, Gang and Su, Min and Zhou, Jianhua and Li, Fei},
  title = {Noninvasive Estimation of Tumor Interstitial Fluid Pressure from Subharmonic Scattering of Ultrasound Contrast Microbubbles},
  journal = {Biosensors},
  year = {2023},
  doi = {10.3390/bios13050528}
}
Kostenlos starten →