EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Deep Implant-Associated Infection in Magnetically Controlled Versus Traditional Growing Rods for Early-Onset Scoliosis: A Narrative Review of Clinical Evidence and Bioengineering Considerations

Michał Grabala, Joanna Suszczyńska, Magdalena Rybaczek, Paweł Grabala

Bioengineering · 21.11.2026 · 13 (12) · 1405

DOI: 10.3390/bioengineering13121405 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 21.11.2026. Verfasst von Michał Grabala und 3 weiteren Autor:innen. Die Arbeit nennt 106 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Magnetically controlled growing rods (MCGRs) avoid routine operative lengthenings required by traditional growing rods (TGRs), but their effect on deep infection remains uncertain. This narrative review integrates clinical infection outcomes with microbiology and bioengineering evidence. A documented revision search used PubMed and Google Scholar, supplemented by scholarly discovery tools; a source-access inventory and MINORS-informed appraisal identify evidentiary limitations. Deep infection was separated from superficial infection, wound complications, and mechanical revision. An MCGR registry reported 33 infected patients among 992 recipients. In one matched cohort of children with severe early-onset scoliosis (EOS) with 44 children per group, reported deep surgical site infection counts were 1 with MCGRs and 11 with TGRs at two years (reported risk ratio, 0.091; 95% confidence interval, 0.012–0.674). This hypothesis-generating observational comparison involved different treatment eras and does not establish a consistent device effect across EOS populations. Retrieval studies document wear but are conditioned on removal; laboratory adhesion, tissue discoloration, and circulating metal do not diagnose clinical infection. Selected constructs may be retained after debridement and antimicrobial treatment, but retention does not establish eradication or equivalence to exchange. Standardized infection definitions, complete treatment histories, and longer comparative follow-up are needed.

Zitieren

APA

Grabala, M., Suszczyńska, J., Rybaczek, M., & Grabala, P. (2026). Deep Implant-Associated Infection in Magnetically Controlled Versus Traditional Growing Rods for Early-Onset Scoliosis: A Narrative Review of Clinical Evidence and Bioengineering Considerations. Bioengineering, 13(12), 1405. https://doi.org/10.3390/bioengineering13121405

Vancouver

Grabala M, Suszczyńska J, Rybaczek M, Grabala P. Deep Implant-Associated Infection in Magnetically Controlled Versus Traditional Growing Rods for Early-Onset Scoliosis: A Narrative Review of Clinical Evidence and Bioengineering Considerations. Bioengineering. 2026;13(12):1405. doi:10.3390/bioengineering13121405

BibTeX
@article{Grabala2026,
  author = {Grabala, Michał and Suszczyńska, Joanna and Rybaczek, Magdalena and Grabala, Paweł},
  title = {Deep Implant-Associated Infection in Magnetically Controlled Versus Traditional Growing Rods for Early-Onset Scoliosis: A Narrative Review of Clinical Evidence and Bioengineering Considerations},
  journal = {Bioengineering},
  year = {2026},
  volume = {13},
  number = {12},
  pages = {1405},
  doi = {10.3390/bioengineering13121405}
}
Kostenlos starten →