Hemodynamic Effects of Renal Artery Orientation and Partial Ostial Coverage After Endovascular Aneurysm Repair: A 4D Flow MRI-Informed Computational Study
Bioengineering · 20.11.2026 · 13 (11) · 1399
DOI: 10.3390/bioengineering13111399 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bioengineering, online erschienen am 20.11.2026. Verfasst von Pengchao Duan und 5 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Objectives: Partial renal ostial coverage by stent-graft fabric during endovascular aneurysm repair (EVAR) may alter local flow. This study used four-dimensional flow magnetic resonance imaging (4D flow MRI) and computational fluid dynamics (CFD) to examine the effects of renal artery orientation and ostial coverage on renal hemodynamics.
Methods
Nine virtual models were constructed from preoperative computed tomography angiography of one patient with an abdominal aortic aneurysm. Left renal artery orientations of 30° upward, horizontal, and 30° downward were combined with nominal linear coverage of 0%, 30%, or 50%. Pulsatile CFD incorporated an MRI-derived inlet waveform, rigid walls, a Carreau viscosity model, and prescribed three-element Windkessel outlet conditions with identical parameters across geometries. CFD results for the horizontal reference model without ostial coverage were compared qualitatively with preoperative 4D flow MRI measurements. Left renal flow fraction was the ratio of cycle-integrated renal outflow to cycle-integrated aortic inflow. Recirculation and wall shear stress-related indices were assessed.
Results
The predicted left renal flow fraction ranged from 16.759% to 17.212% across the nine configurations. Local flow and wall shear patterns differed among configurations. In the uncovered horizontal reference model, the MRI and CFD maps showed broadly similar locations of the main high-velocity regions at the selected systolic phase, with local differences during deceleration. This comparison was qualitative.
Conclusions
The case illustrates that differences in near-ostial flow patterns can coexist with a narrow range of predicted left renal flow fractions under the prescribed conditions. These exploratory findings require further numerical verification and assessment in additional patient anatomies.
Zitieren
APA
Duan, P., Xu, H., Weng, C., Chen, L., Yuan, D., & Zheng, T. (2026). Hemodynamic Effects of Renal Artery Orientation and Partial Ostial Coverage After Endovascular Aneurysm Repair: A 4D Flow MRI-Informed Computational Study. Bioengineering, 13(11), 1399. https://doi.org/10.3390/bioengineering13111399
Vancouver
Duan P, Xu H, Weng C, Chen L, Yuan D, Zheng T. Hemodynamic Effects of Renal Artery Orientation and Partial Ostial Coverage After Endovascular Aneurysm Repair: A 4D Flow MRI-Informed Computational Study. Bioengineering. 2026;13(11):1399. doi:10.3390/bioengineering13111399
BibTeX
@article{Duan2026,
author = {Duan, Pengchao and Xu, Haoyue and Weng, Chengxin and Chen, Luyu and Yuan, Ding and Zheng, Tinghui},
title = {Hemodynamic Effects of Renal Artery Orientation and Partial Ostial Coverage After Endovascular Aneurysm Repair: A 4D Flow MRI-Informed Computational Study},
journal = {Bioengineering},
year = {2026},
volume = {13},
number = {11},
pages = {1399},
doi = {10.3390/bioengineering13111399}
}
EUVIMED