Evaluating Interface Pressure Distribution and Range of Motion in a Mesh-Structure 3D-Printed Lumbar Brace
Bioengineering · 12.08.2026 · 13 (8) · 910
DOI: 10.3390/bioengineering13080910 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bioengineering, online erschienen am 12.08.2026. Verfasst von Hsiang-Chieh Hsu und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physical Therapy and Assistive Technology, National Yang Ming Chiao Tung University, Taipei 112304, Taiwan. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natioanl Science and Technology Council, National Science and Technology Council.
Abstract (Originalsprache)
Conventional spinal braces often suffer from poor fit and low breathability, with most 3D-printing research focusing on scoliosis rather than general support. This study evaluated the biomechanical performance and range of motion (ROM) restriction of a semi-custom 3D-printed brace (3DP) compared with two traditional braces (Medi-H, Medi-S). Sixteen participants performed static standing, flexion, extension, and pick-up tasks while wearing each brace. Interface pressure was recorded across seven anatomical regions, and thoracic (C7–T8) and lumbar (T8–L3) ROM were assessed simultaneously using repeated-measures ANOVA. Results showed that 3DP and Medi-H provided significantly greater overall pressures than Medi-S. During flexion and pick-up tasks, 3DP delivered enhanced support to the middle lumbar region and erector spinae, whereas Medi-H concentrated greater pressure on the spine, particularly during extension. Thoracic and lumbar ROM were restricted comparably across all braces (20–30°). In conclusion, 3DP achieves an equivalent restriction of sagittal motion to traditional braces while offering superior regional support to the upper and middle lumbar regions.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Physical Therapy and Assistive Technology, National Yang Ming Chiao Tung University, Taipei 112304, Taiwan.
ORCID-Profile: Chen-Sheng Chen
Förderung
- Natioanl Science and Technology Council
- National Science and Technology Council
Zitieren
APA
Hsu, H. C., Chen, P. Y., & Chen, C. S. (2026). Evaluating Interface Pressure Distribution and Range of Motion in a Mesh-Structure 3D-Printed Lumbar Brace. Bioengineering, 13(8), 910. https://doi.org/10.3390/bioengineering13080910
Vancouver
Hsu HC, Chen PY, Chen CS. Evaluating Interface Pressure Distribution and Range of Motion in a Mesh-Structure 3D-Printed Lumbar Brace. Bioengineering. 2026;13(8):910. doi:10.3390/bioengineering13080910
BibTeX
@article{Hsu2026,
author = {Hsu, Hsiang-Chieh and Chen, Po-Yin and Chen, Chen-Sheng},
title = {Evaluating Interface Pressure Distribution and Range of Motion in a Mesh-Structure 3D-Printed Lumbar Brace},
journal = {Bioengineering},
year = {2026},
volume = {13},
number = {8},
pages = {910},
doi = {10.3390/bioengineering13080910}
}
EUVIMED