EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Portable TD-OCT Scanner for Dental Use

Tatsuo Shiina, Seiroh Okaneya iD

Bioengineering · 25.06.2026 · 13 (7) · 743

DOI: 10.3390/bioengineering13070743 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 25.06.2026. Verfasst von Tatsuo Shiina und 1 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of Engineering, Chiba University, 1-33 Yayoi-cho, Inage-ku, Chiba-shi 263-8522, Chiba, Japan. Die Arbeit nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A portable OCT scanner usable for dental applications is designed to be compact (B4 notebook size), lightweight, and capable of DC battery operation, enabling chairside use and mobility in home medical care. TD-OCT faces the challenge of a slower measurement speed compared to SD/SS-OCT. In this study, by utilizing multiple rotating reflectors and combining them with a 3 × 3 fiber coupler, a measurement range of over 10 mm and a measurement speed of 1000 scans per second were achieved. Additionally, the dental intraoral probe was designed to ensure high accessibility, allowing free positioning within the oral cavity, with an ingeniously compact optical system. A cylindrical probe with a diameter of 6 mmϕ and a length of 10 mm was developed to enable tooth measurement. This study demonstrated the capability to evaluate attenuation coefficients derived from material properties, starting with measurements of caries and gums.

Schlagwörter

Stichwörter der Autor:innen

OCTcariesdentalgumportableteeth

Autor:innen und Institutionen

  • Graduate School of Engineering, Chiba University, 1-33 Yayoi-cho, Inage-ku, Chiba-shi 263-8522, Chiba, Japan.
  • Phoenix-Dent Co., Ltd., 2649-3 Kamei-cho, Sano-shi 327-0024, Tochigi, Japan.

ORCID-Profile: Seiroh Okaneya

Zitieren

APA

Shiina, T., & Okaneya, S. (2026). Portable TD-OCT Scanner for Dental Use. Bioengineering, 13(7), 743. https://doi.org/10.3390/bioengineering13070743

Vancouver

Shiina T, Okaneya S. Portable TD-OCT Scanner for Dental Use. Bioengineering. 2026;13(7):743. doi:10.3390/bioengineering13070743

BibTeX
@article{Shiina2026,
  author = {Shiina, Tatsuo and Okaneya, Seiroh},
  title = {Portable TD-OCT Scanner for Dental Use},
  journal = {Bioengineering},
  year = {2026},
  volume = {13},
  number = {7},
  pages = {743},
  doi = {10.3390/bioengineering13070743}
}
Kostenlos starten →