The Effects of Multiple Use and Sterilization on the Accuracy of Readings of Different Types of Torque Controllers

Piotr Stendera iD, Jolanta Kostrzewa-Janicka iD

Bioengineering · 23.06.2026 · 13 (7) · 715

DOI: 10.3390/bioengineering13070715 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 23.06.2026. Verfasst von Piotr Stendera und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthetic Dentistry, Medical University of Warsaw, Ul. Binieckiego 6, 02-097 Warsaw, Poland. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Application of the appropriate torque in implant prosthodontics is very important from a clinical perspective. In implant prosthodontics, three basic types of torque wrenches are used. There are few reports in the dental literature on the possibility of controlling the torque generated by the wrench. The aim of this study was to determine the precision of measurements of various types of torque wrenches after the specified number of application and sterilization cycles. The materials consisted of 22 torque wrenches (10 spring-type, 6 toggle-type, and 6 contra-angle-type). The research methodology included measuring the accuracy of torque value indications using the gravity method. The research was conducted for different torque ranges for brand new wrenches and after 50 and 100 sterilization cycles. After each sterilization cycle, the wrenches were used to generate torque several times.

Results

Before sterilization, the spring-type torque controllers were characterized by the lowest accuracy (in the largest range at low torque). The average deviation for contra-angle-type and toggle-type wrenches was smaller. After sterilization cycles, the toggle-type wrenches showed the smallest changes in measurement precision.

Conclusions

The effects of multiple use and sterilization on the accuracy of readings of different types of torque controllers seem limited overall; the greatest stability of measurement accuracy was demonstrated by the toggle-type wrenches, while the spring-type wrenches presented the greatest decrease in measurement precision.

Schlagwörter

Stichwörter der Autor:innen

torque controltorque verificationtorque wrenches

Autor:innen und Institutionen

  • Department of Prosthetic Dentistry, Medical University of Warsaw, Ul. Binieckiego 6, 02-097 Warsaw, Poland.

ORCID-Profile: Piotr Stendera, Jolanta Kostrzewa-Janicka

Zitieren

APA

Stendera, P., & Kostrzewa-Janicka, J. (2026). The Effects of Multiple Use and Sterilization on the Accuracy of Readings of Different Types of Torque Controllers. Bioengineering, 13(7), 715. https://doi.org/10.3390/bioengineering13070715

Vancouver

Stendera P, Kostrzewa-Janicka J. The Effects of Multiple Use and Sterilization on the Accuracy of Readings of Different Types of Torque Controllers. Bioengineering. 2026;13(7):715. doi:10.3390/bioengineering13070715

BibTeX
@article{Stendera2026,
  author = {Stendera, Piotr and Kostrzewa-Janicka, Jolanta},
  title = {The Effects of Multiple Use and Sterilization on the Accuracy of Readings of Different Types of Torque Controllers},
  journal = {Bioengineering},
  year = {2026},
  volume = {13},
  number = {7},
  pages = {715},
  doi = {10.3390/bioengineering13070715}
}
Kostenlos starten →