EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Detection and Measurement of Hypopyon on Slit Lamp Examination Versus Anterior Segment Optical Coherence Tomography

Kamini N Reddy, Folahan Ibukun iD, Kaiyang Huang, Ji Yi, Elesh Jain, Subeesh Kuyyadiyil, Gautam Parmar, Nakul S Shekhawat iD

Bioengineering · 19.05.2026 · 13 (5) · 582

DOI: 10.3390/bioengineering13050582 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 19.05.2026. Verfasst von Kamini N Reddy und 7 weiteren Autor:innen. Beteiligte Einrichtung: Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USA. Die Arbeit nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NEI NIH HHS, KeraLink International, Stephen F Raab and Mariellen Brickley-Raab Rising Professorship in Ophthalmology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To compare hypopyon detection using anterior segment optical coherence tomography (ASOCT) versus slit lamp examination (SLE) in microbial keratitis, and to evaluate intra- and inter-grader agreement for ASOCT hypopyon measurement.

Methods

Two masked graders independently evaluated ASOCT images for hypopyon presence or absence in eyes with microbial keratitis, with disagreements resolved by consensus. A subset of hypopyon eyes underwent triplicate height measurement using two methods (endothelial length and vertical height). The proportion of eyes with hypopyon, Cohen’s kappa, and intraclass correlation coefficients (ICCs) were calculated comparing diagnostic performance of ASOCT versus SLE.

Results

Inter-grader agreement for hypopyon detection on ASOCT was excellent (κ = 0.94; 95% CI 0.84–1.00) and intra-grader agreement was excellent (κ = 0.89–1.00). ASOCT detected hypopyon in 67.1% of eyes versus 57.0% by SLE. Using ASOCT consensus grading as the reference standard, SLE demonstrated a detection proportion of 83.0% (95% CI, 71.4–92.1%). Intra-grader reproducibility was excellent for both endothelial length and vertical height measurements (ICC 0.977–0.996). Inter-grader agreement was good for endothelial length (ICC 0.831) and vertical height (ICC 0.827), though a statistically significant inter-grader bias was identified for vertical height only (Wilcoxon exact p = 0.006).

Conclusions

Among eyes with gradable ASOCT images, ASOCT detected a greater proportion of hypopyon than SLE and demonstrated excellent intra-grader and good inter-grader measurement reproducibility. Endothelial length showed slightly superior inter-grader concordance to vertical height measurement.

Schlagwörter

Stichwörter der Autor:innen

anterior segment optical coherence tomographydiagnostic concordancehypopyoninfectious keratitismicrobial keratitistelemedicine

Autor:innen und Institutionen

  • Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USA.
  • Whiting School of Engineering, Johns Hopkins University, Baltimore, MD 21218, USA.
  • SNC Chitrakoot, Chitrakoot 485334, Madhya Pradesh, India.

ORCID-Profile: Folahan Ibukun, Nakul S Shekhawat

Förderung

  • NEI NIH HHS
  • KeraLink International
  • Stephen F Raab and Mariellen Brickley-Raab Rising Professorship in Ophthalmology
  • National Institutes of Health
  • Stephen F Raab
  • Mariellen Brickley-Raab

Zitieren

APA

Reddy, K. N., Ibukun, F., Huang, K., Yi, J., Jain, E., Kuyyadiyil, S., Parmar, G., & Shekhawat, N. S. (2026). Detection and Measurement of Hypopyon on Slit Lamp Examination Versus Anterior Segment Optical Coherence Tomography. Bioengineering, 13(5), 582. https://doi.org/10.3390/bioengineering13050582

Vancouver

Reddy KN, Ibukun F, Huang K, Yi J, Jain E, Kuyyadiyil S, et al. Detection and Measurement of Hypopyon on Slit Lamp Examination Versus Anterior Segment Optical Coherence Tomography. Bioengineering. 2026;13(5):582. doi:10.3390/bioengineering13050582

BibTeX
@article{Reddy2026,
  author = {Reddy, Kamini N and Ibukun, Folahan and Huang, Kaiyang and Yi, Ji and Jain, Elesh and Kuyyadiyil, Subeesh and Parmar, Gautam and Shekhawat, Nakul S},
  title = {Detection and Measurement of Hypopyon on Slit Lamp Examination Versus Anterior Segment Optical Coherence Tomography},
  journal = {Bioengineering},
  year = {2026},
  volume = {13},
  number = {5},
  pages = {582},
  doi = {10.3390/bioengineering13050582}
}
Kostenlos starten →