Numerical Simulation and Optimization of Drug-Coated Balloon Inflation for Vascular Stenosis
Bioengineering · 05.03.2026 · 13 (3) · 301
DOI: 10.3390/bioengineering13030301 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bioengineering, online erschienen am 05.03.2026. Verfasst von Chenzhao Zhang und 2 weiteren Autor:innen. Beteiligte Einrichtung: College of Biomedical Engineering, Fudan University, Shanghai 200433, China. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
This study establishes a numerical model for the implantation of drug-coated balloons (DCBs) in blood vessels, in order to quantify drug (paclitaxel) transfer and diffusion within stenotic vessels and provide a reference for clinical surgical plans.
Objective
To study the change in paclitaxel concentration over time in the blood vessel after the implantation of a DCB in vessels with different degrees of stenosis, and thereby determine the optimal balloon dilation time.
Method
Using finite element modeling and numerical simulation techniques, a model was established to study the rules of paclitaxel concentration change over time in vessels with different degrees of stenosis.
Results
Based on the simulation prediction, the longer the balloon dilation time during implantation, the more paclitaxel enters the vessel wall. After the balloon is withdrawn, the paclitaxel gradually diffuses evenly throughout the vessel, and the paclitaxel concentration gradually decreases over time.
Conclusion
Under the simulation conditions, the optimal balloon dilation times for vessels with stenosis rates of 10%, 30%, and 50% should be 20 s, 60 s, and 80 s, corresponding effective duration of 6 weeks, 6 weeks and 4 weeks, respectively.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- College of Biomedical Engineering, Fudan University, Shanghai 200433, China.
- Yiwu Research Institute, Fudan University, Yiwu 322000, China.
ORCID-Profile: Shengzhang Wang
Zitieren
APA
Zhang, C., Zhang, Y., & Wang, S. (2026). Numerical Simulation and Optimization of Drug-Coated Balloon Inflation for Vascular Stenosis. Bioengineering, 13(3), 301. https://doi.org/10.3390/bioengineering13030301
Vancouver
Zhang C, Zhang Y, Wang S. Numerical Simulation and Optimization of Drug-Coated Balloon Inflation for Vascular Stenosis. Bioengineering. 2026;13(3):301. doi:10.3390/bioengineering13030301
BibTeX
@article{Zhang2026,
author = {Zhang, Chenzhao and Zhang, Yuanyuan and Wang, Shengzhang},
title = {Numerical Simulation and Optimization of Drug-Coated Balloon Inflation for Vascular Stenosis},
journal = {Bioengineering},
year = {2026},
volume = {13},
number = {3},
pages = {301},
doi = {10.3390/bioengineering13030301}
}
EUVIMED