EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Correlation Analysis of Suture Anchor Pull-Out Strength with Cortical Bone Thickness and Cancellous Bone Density on a Finite Element Model

Jung Ho Kim, Jeon Jong Hyeok, Jae Hyun Woo iD, Sung Min Kim

Bioengineering · 11.08.2025 · 12 (8) · 863

DOI: 10.3390/bioengineering12080863 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 11.08.2025. Verfasst von Jung Ho Kim und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Regulatory Science for Medical Device, Dongguk University, Seoul 04620, Republic of Korea. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The Ministry of Food and Drug, Ministry of Food and Drug Safety of South Korea.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study aimed to assess, using finite element analysis (FEA), the mechanical effects of cortical bone thickness and cancellous bone density on the pull-out strength of suture anchors. A PEEK anchor was modeled and embedded in synthetic bone blocks with cortical thicknesses ranging from 1 to 5 mm and cancellous densities of 10 PCF, 20 PCF, and 30 PCF. Axial tensile loading simulations were conducted for all combinations, and selected cases were validated through experimental pull-out tests using commercial synthetic bone, demonstrating agreement within ±6%. Both cortical thickness and cancellous density were found to enhance pull-out resistance, though the magnitude and pattern varied with density. At 10 PCF, pull-out strength increased linearly with cortical thickness. At 20 PCF, substantial gains were observed between 2 and 4 mm, followed by a plateau. At 30 PCF, most of the increase was confined between 2 and 3 mm, with minimal improvement thereafter. These findings suggest that fixation strategies should be adapted on the basis of bone quality and provide biomechanical insights to inform patient-specific implant design and surgical planning.

Schlagwörter

Stichwörter der Autor:innen

Finite Element Method (FEM)cancellous bone densitycortical bone thicknesssuture anchor

Autor:innen und Institutionen

  • Department of Regulatory Science for Medical Device, Dongguk University, Seoul 04620, Republic of Korea.
  • Department of Medical and Healthcare, Dongguk University, Seoul 04620, Republic of Korea.
  • Department of Biomedical Engineering, College of Life Science and Biotechnology, Dongguk University, Seoul 04620, Republic of Korea.

ORCID-Profile: Jae Hyun Woo

Förderung

  • The Ministry of Food and Drug
  • Ministry of Food and Drug Safety of South Korea

Zitieren

APA

Kim, J. H., Hyeok, J. J., Woo, J. H., & Kim, S. M. (2025). Correlation Analysis of Suture Anchor Pull-Out Strength with Cortical Bone Thickness and Cancellous Bone Density on a Finite Element Model. Bioengineering, 12(8), 863. https://doi.org/10.3390/bioengineering12080863

Vancouver

Kim JH, Hyeok JJ, Woo JH, Kim SM. Correlation Analysis of Suture Anchor Pull-Out Strength with Cortical Bone Thickness and Cancellous Bone Density on a Finite Element Model. Bioengineering. 2025;12(8):863. doi:10.3390/bioengineering12080863

BibTeX
@article{Kim2025,
  author = {Kim, Jung Ho and Hyeok, Jeon Jong and Woo, Jae Hyun and Kim, Sung Min},
  title = {Correlation Analysis of Suture Anchor Pull-Out Strength with Cortical Bone Thickness and Cancellous Bone Density on a Finite Element Model},
  journal = {Bioengineering},
  year = {2025},
  volume = {12},
  number = {8},
  pages = {863},
  doi = {10.3390/bioengineering12080863}
}