High-Throughput Microfluidic Electroporation (HTME): A Scalable, 384-Well Platform for Multiplexed Cell Engineering
Bioengineering · 22.07.2025 · 12 (8) · 788
DOI: 10.3390/bioengineering12080788 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bioengineering, online erschienen am 22.07.2025. Verfasst von William R Gaillard und 12 weiteren Autor:innen. Beteiligte Einrichtung: DOE Joint BioEnergy Institute, Emeryville, CA 94608, USA. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: United States Department of Energy, Joint BioEnergy Institute, U.S. Department of Energy, Office of Science, Biological and Environmental Research Program.
Abstract (Originalsprache)
Electroporation-mediated gene delivery is a cornerstone of synthetic biology, offering several advantages over other methods: higher efficiencies, broader applicability, and simpler sample preparation. Yet, electroporation protocols are often challenging to integrate into highly multiplexed workflows, owing to limitations in their scalability and tunability. These challenges ultimately increase the time and cost per transformation. As a result, rapidly screening genetic libraries, exploring combinatorial designs, or optimizing electroporation parameters requires extensive iterations, consuming large quantities of expensive custom-made DNA and cell lines or primary cells. To address these limitations, we have developed a High-Throughput Microfluidic Electroporation (HTME) platform that includes a 384-well electroporation plate (E-Plate) and control electronics capable of rapidly electroporating all wells in under a minute with individual control of each well. Fabricated using scalable and cost-effective printed-circuit-board (PCB) technology, the E-Plate significantly reduces consumable costs and reagent consumption by operating on nano to microliter volumes. Furthermore, individually addressable wells facilitate rapid exploration of large sets of experimental conditions to optimize electroporation for different cell types and plasmid concentrations/types. Use of the standard 384-well footprint makes the platform easily integrable into automated workflows, thereby enabling end-to-end automation. We demonstrate transformation of E. coli with pUC19 to validate the HTME’s core functionality, achieving at least a single colony forming unit in more than 99% of wells and confirming the platform’s ability to rapidly perform hundreds of electroporations with customizable conditions. This work highlights the HTME’s potential to significantly accelerate synthetic biology Design-Build-Test-Learn (DBTL) cycles by mitigating the transformation/transfection bottleneck.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- DOE Joint BioEnergy Institute, Emeryville, CA 94608, USA.
- Sandia National Laboratories, Livermore, CA 94550, USA.
- Biological Systems & Engineering Division, Lawrence Berkeley National Laboratory, Berkeley, CA 94720, USA.
- Engineering Directorate, Lawrence Livermore National Laboratory, Livermore, CA 94550, USA.
- Molecular Biophysics and Integrated Bioimaging Division, Lawrence Berkeley National Laboratory, Berkeley, CA 94720, USA.
- Department of Bioengineering, University of California Berkeley, Berkeley, CA 94720, USA.
ORCID-Profile: David N Carruthers, Kshitiz Gupta, Hector Garcia Martin, Nathan J Hillson
Förderung
- United States Department of Energy
- Joint BioEnergy Institute, U.S. Department of Energy, Office of Science, Biological and Environmental Research Program
Zitieren
APA
Gaillard, W. R., Sustarich, J., Li, Y., Carruthers, D. N., Gupta, K., Liang, Y., Kuo, R., Tan, S., Yoder, S., Adams, P. D., Garcia Martin, H., Hillson, N. J., & Singh, A. K. (2025). High-Throughput Microfluidic Electroporation (HTME): A Scalable, 384-Well Platform for Multiplexed Cell Engineering. Bioengineering, 12(8), 788. https://doi.org/10.3390/bioengineering12080788
Vancouver
Gaillard WR, Sustarich J, Li Y, Carruthers DN, Gupta K, Liang Y, et al. High-Throughput Microfluidic Electroporation (HTME): A Scalable, 384-Well Platform for Multiplexed Cell Engineering. Bioengineering. 2025;12(8):788. doi:10.3390/bioengineering12080788
BibTeX
@article{Gaillard2025,
author = {Gaillard, William R and Sustarich, Jess and Li, Yuerong and Carruthers, David N and Gupta, Kshitiz and Liang, Yan and Kuo, Rita and Tan, Stephen and Yoder, Sam and Adams, Paul D and Garcia Martin, Hector and Hillson, Nathan J and Singh, Anup K},
title = {High-Throughput Microfluidic Electroporation (HTME): A Scalable, 384-Well Platform for Multiplexed Cell Engineering},
journal = {Bioengineering},
year = {2025},
volume = {12},
number = {8},
pages = {788},
doi = {10.3390/bioengineering12080788}
}
EUVIMED