EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Optimization of Tibial Stem Geometry in Total Knee Arthroplasty Using Design of Experiments: A Finite Element Analysis

Hyun Hee Lee iD, Hyoung-Taek Hong, Jong-Keun Kim, Yong-Gon Koh, Kwan Kyu Park iD, Kyoung-Tak Kang

Bioengineering · 11.02.2025 · 12 (2) · 172

DOI: 10.3390/bioengineering12020172 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 11.02.2025. Verfasst von Hyun Hee Lee und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, International St. Mary's Hospital, Incheon 22711, Republic of Korea. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The stability of the tibial component in Total Knee Arthroplasty (TKA) is critical to preventing aseptic loosening, a major cause of implant failure. However, existing tibial stem designs often lead to stress shielding and bone resorption, highlighting the need for further optimization. This study addresses these challenges by employing the Design of Experiments (DOE) methodology, specifically utilizing a full factorial design approach combined with finite element analysis (FEA), to optimize the geometry of the tibial stem. The material properties of the cortical and cancellous bone, as well as the tibial tray, were assigned based on values from the literature, representing their elastic moduli and Poisson’s ratios. For boundary conditions, the distal end of the tibia was fully constrained to simulate realistic load transfer, while compressive loads representative of walking and daily activities were applied to the tibial base. Key design parameters, including stem diameter, length, mediolateral ratio (M/L ratio), and wing angle, were systematically analyzed. The results identified stem diameter and length as the most influential factors in improving biomechanical performance, while the wing angle showed minimal impact. The optimized design, featuring a stem diameter of 12 mm, length of 40 mm, M/L ratio of 0.61, and a wing angle of 60°, demonstrated significant reductions in stress shielding and aseptic loosening compared to conventional models. These findings provide valuable insights into enhancing the long-term success of TKA implants by balancing implant stability and minimizing bone resection.

Schlagwörter

Stichwörter der Autor:innen

design of experimentsfinite element analysistibial stem geometry

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, International St. Mary's Hospital, Incheon 22711, Republic of Korea.
  • Skyve R&D LAB, Seoul 07217, Republic of Korea.
  • Department of Orthopaedic Surgery, Heung-K Hospital, Siheung 14999, Republic of Korea.
  • Joint Reconstruction Center, Department of Orthopaedic Surgery, Yonsei Sarang Hospital, Seoul 06702, Republic of Korea.
  • Department of Orthopedic Surgery, Severance Hospital, Yonsei University College of Medicine, Seoul 03722, Republic of Korea.
  • Department of Mechanical Engineering, Yonsei University, Seoul 03722, Republic of Korea.

ORCID-Profile: Hyun Hee Lee, Kwan Kyu Park

Zitieren

APA

Lee, H. H., Hong, H. T., Kim, J. K., Koh, Y. G., Park, K. K., & Kang, K. T. (2025). Optimization of Tibial Stem Geometry in Total Knee Arthroplasty Using Design of Experiments: A Finite Element Analysis. Bioengineering, 12(2), 172. https://doi.org/10.3390/bioengineering12020172

Vancouver

Lee HH, Hong HT, Kim JK, Koh YG, Park KK, Kang KT. Optimization of Tibial Stem Geometry in Total Knee Arthroplasty Using Design of Experiments: A Finite Element Analysis. Bioengineering. 2025;12(2):172. doi:10.3390/bioengineering12020172

BibTeX
@article{Lee2025,
  author = {Lee, Hyun Hee and Hong, Hyoung-Taek and Kim, Jong-Keun and Koh, Yong-Gon and Park, Kwan Kyu and Kang, Kyoung-Tak},
  title = {Optimization of Tibial Stem Geometry in Total Knee Arthroplasty Using Design of Experiments: A Finite Element Analysis},
  journal = {Bioengineering},
  year = {2025},
  volume = {12},
  number = {2},
  pages = {172},
  doi = {10.3390/bioengineering12020172}
}