EUVIMEDEUVIMED entdecken
StudienprotokollOpen Access · CC BY 4.0

Comparison of One-Drill Protocol to Sequential Drilling In Vitro and In Vivo

Sihana Rugova, Marcus Abboud iD

Bioengineering · 09.01.2025 · 12 (1) · 51

DOI: 10.3390/bioengineering12010051 ↗

Kurz gefasst

Studienprotokoll in der Zeitschrift Bioengineering, online erschienen am 09.01.2025. Verfasst von Sihana Rugova und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral Biology and Pathology, Stony Brook University, Stony Brook, NY 11794, USA. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: This research received NIH funding, award 74009, active from December 2015 to February 2017., NIH funding.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study compares the heat generated during bone drilling using different protocols and implant systems, first in vitro and then in vivo with an animal model. In the experimental phase, thermal data were collected using an infrared camera while preparing implant beds in bone similes. The heat generated by a one-drill protocol with a new-generation drill bit and the Straumann BLT sequential drilling protocol was evaluated. The experimental study was then replicated in an animal model to assess the impact of these protocols on early osseointegration, measured by bone-to-implant contact (BIC) at three weeks post-surgery for Straumann BLT SLActive and Medentika Quattrocone implants. The results showed the BLT sequential protocol generated significantly more heat during drilling in bone similes compared to the new-generation drill bit. In the animal model, a histological analysis revealed a trend favoring shorter drilling protocols, with reduced drilling times and a potential advantage for osseointegration, though the BIC differences were not statistically significant. These findings suggest that minimizing the number of drilling steps and thermal stress may enhance osseointegration more effectively than advanced implant surface treatments. This aligns with emerging views on the importance of optimized drilling protocols and designs to reduce heat generation and better preserve surrounding bone structure.

Schlagwörter

Stichwörter der Autor:innen

BLTSLActivebone-to-implant contactdental implantdrilling protocolsheat generationimplant bed preparationimplant drillsosseointegrationosteotomystraumannsurgical drill bitsthermal trauma

Autor:innen und Institutionen

  • Department of Oral Biology and Pathology, Stony Brook University, Stony Brook, NY 11794, USA.
  • School of Engineering, Stony Brook University, Stony Brook, NY 11794, USA.

ORCID-Profile: Marcus Abboud

Förderung

  • This research received NIH funding, award 74009, active from December 2015 to February 2017.
  • NIH funding

Zitieren

APA

Rugova, S., & Abboud, M. (2025). Comparison of One-Drill Protocol to Sequential Drilling In Vitro and In Vivo. Bioengineering, 12(1), 51. https://doi.org/10.3390/bioengineering12010051

Vancouver

Rugova S, Abboud M. Comparison of One-Drill Protocol to Sequential Drilling In Vitro and In Vivo. Bioengineering. 2025;12(1):51. doi:10.3390/bioengineering12010051

BibTeX
@article{Rugova2025,
  author = {Rugova, Sihana and Abboud, Marcus},
  title = {Comparison of One-Drill Protocol to Sequential Drilling In Vitro and In Vivo},
  journal = {Bioengineering},
  year = {2025},
  volume = {12},
  number = {1},
  pages = {51},
  doi = {10.3390/bioengineering12010051}
}
Kostenlos starten →