Two-Dimensional Mammography Imaging Techniques for Screening Women with Silicone Breast Implants: A Pilot Phantom Study

Isabelle Fitton iD, Virginia Tsapaki iD, Jonathan Zerbib, Antoine Decoux, Amit Kumar iD, Aude Stembert, Françoise Malchair, Claire Van Ngoc Ty iD, Laure Fournier

Bioengineering · 31.08.2024 · 11 (9) · 884

DOI: 10.3390/bioengineering11090884 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 31.08.2024. Verfasst von Isabelle Fitton und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, AP-HP, Hôpital Européen Georges Pompidou, 75015 Paris, France. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 54 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study aimed to evaluate the impact of three two-dimensional (2D) mammographic acquisition techniques on image quality and radiation dose in the presence of silicone breast implants (BIs). Then, we propose and validate a new International Atomic Energy Agency (IAEA) phantom to reproduce these techniques. Images were acquired on a single Hologic Selenia Dimensions® unit. The mammography of the left breast of a single clinical case was included. Three methods of image acquisition were identified. They were based on misused, recommended, and reference settings. In the clinical case, image criteria scoring and the signal-to-noise ratio on breast tissue (SNRBT) were determined for two 2D projections and compared between the three techniques. The phantom study first compared the reference and misused settings by varying the AEC sensor position and, second, the recommended settings with a reduced current-time product (mAs) setting that was 13% lower. The signal-difference-to-noise ratio (SDNR) and detectability indexes at 0.1 mm (d’ 0.1 mm) and 0.25 mm (d’ 0.25 mm) were automatically quantified using ATIA software. Average glandular dose (AGD) values were collected for each acquisition. A statistical analysis was performed using Kruskal–Wallis and corrected Dunn tests (p < 0.05). The SNRBT was 2.6 times lower and the AGD was −18% lower with the reference settings compared to the recommended settings. The SNRBT values increased by +98% with the misused compared to the recommended settings. The AGD increased by +79% with the misused settings versus the recommended settings. The median values of the reference settings were 5.8 (IQR 5.7–5.9), 1.2 (IQR 0.0), 7.0 (IQR 6.8–7.2) and 1.2 (IQR 0.0) mGy and were significantly lower than those of the misused settings (p < 0.03): 7.9 (IQR 6.1–9.7), 1.6 (IQR 1.3–1.9), 9.2 (IQR 7.5–10.9) and 2.2 (IQR 1.4–3.0) mGy for the SDNR, d’ 0.1 mm, d’ 0.25 mm and the AGD, respectively. A comparison of the recommended and reduced settings showed a reduction of −6.1 ± 0.6% (p = 0.83), −7.7 ± 0.0% (p = 0.18), −6.4 ± 0.6% (p = 0.19) and −13.3 ± 1.1% (p = 0.53) for the SDNR, d’ 0.1 mm, d’ 0.25 mm and the AGD, respectively. This study showed that the IAEA phantom could be used to reproduce the three techniques for acquiring 2D mammography images in the presence of breast implants for raising awareness and for educational purposes. It could also be used to evaluate and optimize the manufacturer’s recommended settings.

Schlagwörter

Stichwörter der Autor:innen

breastbreast implanteducationionizing radiationmammographyquality control

Autor:innen und Institutionen

  • Department of Radiology, AP-HP, Hôpital Européen Georges Pompidou, 75015 Paris, France.
  • Division of Human Health, Department of Nuclear Sciences and Applications, International Atomic Energy Agency, 1220 Vienna, Austria.
  • Paris Cardiovascular Research Center, Institut National de la Santé et de la Recherche Médicale Unité 970, 75015 Paris, France.
  • OKOMERA, iPEPS, The Healthtech Hub, 75013 Paris, CEDEX 13, France.
  • ZEPHYRA, 4000 Liège, Belgium.
  • Department of Radiology, PARCC UMRS 970, INSERM, Hôpital Européen Georges Pompidou, Université Paris Cité, AP-HP, 75015 Paris, France.

ORCID-Profile: Isabelle Fitton, Virginia Tsapaki, Amit Kumar, Claire Van Ngoc Ty

Zitieren

APA

Fitton, I., Tsapaki, V., Zerbib, J., Decoux, A., Kumar, A., Stembert, A., Malchair, F., Van Ngoc Ty, C., & Fournier, L. (2024). Two-Dimensional Mammography Imaging Techniques for Screening Women with Silicone Breast Implants: A Pilot Phantom Study. Bioengineering, 11(9), 884. https://doi.org/10.3390/bioengineering11090884

Vancouver

Fitton I, Tsapaki V, Zerbib J, Decoux A, Kumar A, Stembert A, et al. Two-Dimensional Mammography Imaging Techniques for Screening Women with Silicone Breast Implants: A Pilot Phantom Study. Bioengineering. 2024;11(9):884. doi:10.3390/bioengineering11090884

BibTeX
@article{Fitton2024,
  author = {Fitton, Isabelle and Tsapaki, Virginia and Zerbib, Jonathan and Decoux, Antoine and Kumar, Amit and Stembert, Aude and Malchair, Françoise and Van Ngoc Ty, Claire and Fournier, Laure},
  title = {Two-Dimensional Mammography Imaging Techniques for Screening Women with Silicone Breast Implants: A Pilot Phantom Study},
  journal = {Bioengineering},
  year = {2024},
  volume = {11},
  number = {9},
  pages = {884},
  doi = {10.3390/bioengineering11090884}
}
Kostenlos starten →