Active Claw-Shaped Dry Electrodes for EEG Measurement in Hair Areas
Bioengineering · 13.03.2024 · 11 (3) · 276
DOI: 10.3390/bioengineering11030276 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bioengineering, online erschienen am 13.03.2024. Verfasst von Zaihao Wang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Center for Intelligent Medical Electronics, School of Information Science and Technology, Fudan University, Shanghai 200433, China. Die Arbeit wurde bisher 20-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Greater Bay Area Institute of Precision Medicine (Guangzhou).
Abstract (Originalsprache)
EEG, which can provide brain alteration information via recording the electrical activity of neurons in the cerebral cortex, has been widely used in neurophysiology. However, conventional wet electrodes in EEG monitoring typically suffer from inherent limitations, including the requirement of skin pretreatment, the risk of superficial skin infections, and signal performance deterioration that may occur over time due to the air drying of the conductive gel. Although the emergence of dry electrodes has overcome these shortcomings, their electrode–skin contact impedance is significantly high and unstable, especially in hair-covered areas. To address the above problems, an active claw-shaped dry electrode is designed, moving from electrode morphological design, slurry preparation, and coating to active electrode circuit design. The active claw-shaped dry electrode, which consists of a claw-shaped electrode and active electrode circuit, is dedicated to offering a flexible solution for elevating electrode fittings on the scalp in hair-covered areas, reducing electrode–skin contact impedance and thus improving the quality of the acquired EEG signal. The performance of the proposed electrodes was verified by impedance, active electrode circuit, eyes open-closed, steady-state visually evoked potential (SSVEP), and anti-interference tests, based on EEG signal acquisition. Experimental results show that the proposed claw-shaped electrodes (without active circuit) can offer a better fit between the scalp and electrodes, with a low electrode–skin contact impedance (18.62 KΩ@1 Hz in the hairless region and 122.15 KΩ@1 Hz in the hair-covered region). In addition, with the active circuit, the signal-to-noise ratio (SNR) of the acquiring EEG signal was improved and power frequency interference was restrained, therefore, the proposed electrodes can yield an EEG signal quality comparable to wet electrodes.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center for Intelligent Medical Electronics, School of Information Science and Technology, Fudan University, Shanghai 200433, China.
- Center for Intelligent Medical Equipment and Devices, Institute for Innovative Medical Devices, School of Biomedical Engineering, Division of Life Sciences and Medicine, University of Science and Technology of China, He…
- Suzhou Institute for Advanced Research, University of Science and Technology of China, Suzhou 215123, China.
- School of Biomedical Engineering, The University of Sydney, Sydney, NSW 2006, Australia.
- Human Phenome Institute, Fudan University, Shanghai 201203, China.
ORCID-Profile: Hongyu Chen
Förderung
- Greater Bay Area Institute of Precision Medicine (Guangzhou)
Zitieren
APA
Wang, Z., Ding, Y., Yuan, W., Chen, H., Chen, W., & Chen, C. (2024). Active Claw-Shaped Dry Electrodes for EEG Measurement in Hair Areas. Bioengineering, 11(3), 276. https://doi.org/10.3390/bioengineering11030276
Vancouver
Wang Z, Ding Y, Yuan W, Chen H, Chen W, Chen C. Active Claw-Shaped Dry Electrodes for EEG Measurement in Hair Areas. Bioengineering. 2024;11(3):276. doi:10.3390/bioengineering11030276
BibTeX
@article{Wang2024,
author = {Wang, Zaihao and Ding, Yuhao and Yuan, Wei and Chen, Hongyu and Chen, Wei and Chen, Chen},
title = {Active Claw-Shaped Dry Electrodes for EEG Measurement in Hair Areas},
journal = {Bioengineering},
year = {2024},
volume = {11},
number = {3},
pages = {276},
doi = {10.3390/bioengineering11030276}
}
EUVIMED