Manual Insertion of Cochlear Implant Electrodes Versus Robot-Assisted Insertion and Analysis by Micro-CT: A Temporal Bone Study
Audiology Research · 26.03.2026 · 16 (2) · 51
DOI: 10.3390/audiolres16020051 ↗Kurz gefasst
Fachartikel in der Zeitschrift Audiology Research, online erschienen am 26.03.2026. Verfasst von Alexandre Karkas und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otorhinolaryngology-Head & Neck Surgery, University Medical Center of Saint-Etienne, 42000 Saint-Etienne, France. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Objectives: Atraumatic electrode array insertion should be targeted in cochlear implantation. Robotic insertion is used in many centers worldwide. Our objective was to evaluate manual electrode placement and robot-assisted placement using RobOtol® on human temporal bones (TBs), in terms of endocochlear trauma and completion of insertion.
Methods
Sixteen TBs originating from eight bodies were implanted with Medel-FLEX24 electrodes through the round window. The right TB was implanted manually, while the left TB of the same body was implanted using RobOtol® for electrode insertion. Results were analyzed through micro-computed tomography imaging. No statistical analysis was used, given the small sample size; a descriptive interpretation of micro-CT scans was rather preferred.
Results
In the “manual group”, there were two cases (25%) of insertion trauma: elevation of basilar membrane at basal turn (Eshraghi-stage-1). In the “robotic group”, there were two cases (25%) of insertion trauma: one case of elevation of basilar membrane at the middle turn (Eshraghi-stage-1) and one case of dislocation of all electrodes in scala vestibuli (Eshraghi-stage-3). There were six cases (75%) of incomplete insertion in the “manual group” and four cases (50%) of incomplete insertion in the “robotic group”.
Conclusions
Both techniques of electrode placement yielded fairly similar results, in terms of endocochlear trauma and completion of insertion. New larger-scale cadaveric and clinical studies are needed to determine the possible benefit of robot-assisted electrode insertion in cochlear implantation.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otorhinolaryngology-Head & Neck Surgery, University Medical Center of Saint-Etienne, 42000 Saint-Etienne, France.
- Medical School, Jean Monnet University, 42270 Saint-Priest-en-Jarez, France.
- Research Laboratory SAINBIOSE Inserm U1059, 42000 Saint-Etienne, France.
- Anatomical Laboratory, Medical School, Jean Monnet University, 42000 Saint-Etienne, France.
- Department of Public Health and Statistics, University Medical Center of Saint-Etienne, 42000 Saint-Etienne, France.
- Department of Rheumatology, University Medical Center of Saint-Etienne, 42270 Saint-Priest-en-Jarez, France.
ORCID-Profile: Hubert Marotte
Zitieren
APA
Karkas, A., Arnold, C., Lelonge, Y., Laroche, N., Tinquaut, F., Bergandi, F., Marotte, H., & Daouda, K. (2026). Manual Insertion of Cochlear Implant Electrodes Versus Robot-Assisted Insertion and Analysis by Micro-CT: A Temporal Bone Study. Audiology Research, 16(2), 51. https://doi.org/10.3390/audiolres16020051
Vancouver
Karkas A, Arnold C, Lelonge Y, Laroche N, Tinquaut F, Bergandi F, et al. Manual Insertion of Cochlear Implant Electrodes Versus Robot-Assisted Insertion and Analysis by Micro-CT: A Temporal Bone Study. Audiology Research. 2026;16(2):51. doi:10.3390/audiolres16020051
BibTeX
@article{Karkas2026,
author = {Karkas, Alexandre and Arnold, Clément and Lelonge, Yann and Laroche, Norbert and Tinquaut, Fabien and Bergandi, Florian and Marotte, Hubert and Daouda, Kelly},
title = {Manual Insertion of Cochlear Implant Electrodes Versus Robot-Assisted Insertion and Analysis by Micro-CT: A Temporal Bone Study},
journal = {Audiology Research},
year = {2026},
volume = {16},
number = {2},
pages = {51},
doi = {10.3390/audiolres16020051}
}