Sub-Micron Particles Recovered from the Irrigant During Ultrasonic Instrumentation of Dental Implants: An In Vitro Study

Wonsup Lee, Jun-Beom Park

Applied Sciences · 16.11.2026 · 16 (22) · 11332

DOI: 10.3390/app162211332 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Applied Sciences, online erschienen am 16.11.2026. Verfasst von Wonsup Lee und 1 weiteren Autor:innen. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Research Fund of Seoul St. Mary’s Hospital, The Catholic University of Kore.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Ultrasonic instrumentation is routine in dental implant maintenance, and the particulate debris it generates has been proposed as a modifier of the peri-implant environment. How much reaches the irrigant, and how much arises from the equipment rather than the implant, is unquantified. This in vitro study measured sub-micron particles recovered from the irrigant after ultrasonic instrumentation of dental implants. In a 2 × 2 factorial design, tip material (stainless steel vs. plastic) and condition (contact vs. non-contact) were fixed factors (n = 9 per group; 36 specimens). Implants were instrumented for 75 s at maximum power under a constant 0.25 N force, and the irrigant was analyzed by nanoparticle tracking analysis. Mean particle size ranged from 245.7 ± 74.2 nm (steel/non-contact) to 349.0 ± 201.0 nm (steel/contact), with no significant effect of tip material (p = 0.348), condition (p = 0.227) or their interaction (p = 0.216); all partial η2 values were ≤ 0.13. Concentrations of 107–108 particles/mL were recovered even when the tip never contacted the implant. Because nanoparticle tracking analysis cannot establish elemental composition, the particles are reported as recovered from the irrigant rather than as implant-derived; the background load they represent must be controlled in future work.

Förderung

  • Research Fund of Seoul St. Mary’s Hospital, The Catholic University of Kore

Zitieren

APA

Lee, W., & Park, J. (2026). Sub-Micron Particles Recovered from the Irrigant During Ultrasonic Instrumentation of Dental Implants: An In Vitro Study. Applied Sciences, 16(22), 11332. https://doi.org/10.3390/app162211332

Vancouver

Lee W, Park J. Sub-Micron Particles Recovered from the Irrigant During Ultrasonic Instrumentation of Dental Implants: An In Vitro Study. Applied Sciences. 2026;16(22):11332. doi:10.3390/app162211332

BibTeX
@article{Lee2026,
  author = {Lee, Wonsup and Park, Jun-Beom},
  title = {Sub-Micron Particles Recovered from the Irrigant During Ultrasonic Instrumentation of Dental Implants: An In Vitro Study},
  journal = {Applied Sciences},
  year = {2026},
  volume = {16},
  number = {22},
  pages = {11332},
  doi = {10.3390/app162211332}
}
Kostenlos starten →