The Effect of Scatter Radiation on Image Resolution in Gridless Portable X-Ray Imaging: A Monte Carlo Study

Ilias Anagnostou, Panagiotis Liaparinos, Christos Michail, Ioannis Valais, George Fountos, Ioannis Kandarakis, Nektarios Kalyvas

Applied Sciences · 25.03.2026 · 16 (7) · 3152

DOI: 10.3390/app16073152 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Applied Sciences, online erschienen am 25.03.2026. Verfasst von Ilias Anagnostou und 6 weiteren Autor:innen. Die Arbeit nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of West Attica.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In X-ray imaging, tissue scattering is an important factor that degrades image clarity, especially using a portable gridless X-ray imaging device. This study focuses on using Monte Carlo simulation to quantify the effect of scatter radiation on image resolution, by analyzing the point spread function (PSF) and the corresponding modulation transfer function (MTF). Lateral energy absorption profiles in tissue and a cesium iodide (CsI) scintillator were calculated at different X-ray tube voltages (70–90 kV) and filter configurations. Results showed that 85.7% of the total scattered radiation is concentrated at a distance of 4 cm from the central axis for the tissue and 67.37% for the CsI scintillator. The MTF remained high at low spatial frequencies (23% at 0.04 cycles/cm) but dropped at mid frequencies (0.015–0.025 at 0.3–0.6 cycles/cm) and was almost zero at high frequencies (0.004 at 0.8 cycles/cm), indicating loss of detail due to scattering. Increasing the thickness of the filter or adding a copper (Cu) filter reduced the contrast at low spatial frequencies (from 23% to 21%). The study quantitatively investigated the MTF degradation in portable X-ray imaging devices without grid, due to scatter. These results may aid in the development of scatter correction algorithms to improve image quality without the need for an anti-scatter grid.

Förderung

  • University of West Attica

Zitieren

APA

Anagnostou, I., Liaparinos, P., Michail, C., Valais, I., Fountos, G., Kandarakis, I., & Kalyvas, N. (2026). The Effect of Scatter Radiation on Image Resolution in Gridless Portable X-Ray Imaging: A Monte Carlo Study. Applied Sciences, 16(7), 3152. https://doi.org/10.3390/app16073152

Vancouver

Anagnostou I, Liaparinos P, Michail C, Valais I, Fountos G, Kandarakis I, et al. The Effect of Scatter Radiation on Image Resolution in Gridless Portable X-Ray Imaging: A Monte Carlo Study. Applied Sciences. 2026;16(7):3152. doi:10.3390/app16073152

BibTeX
@article{Anagnostou2026,
  author = {Anagnostou, Ilias and Liaparinos, Panagiotis and Michail, Christos and Valais, Ioannis and Fountos, George and Kandarakis, Ioannis and Kalyvas, Nektarios},
  title = {The Effect of Scatter Radiation on Image Resolution in Gridless Portable X-Ray Imaging: A Monte Carlo Study},
  journal = {Applied Sciences},
  year = {2026},
  volume = {16},
  number = {7},
  pages = {3152},
  doi = {10.3390/app16073152}
}
Kostenlos starten →