Characterization of the Physical Image Quality of a Clinical Photon-Counting Computed Tomography Scanner Across Multiple Acquisition and Reconstruction Settings

Patrizio Barca, Luigi Masturzo, Luca De Masi, Antonio Traino, Filippo Cademartiri, Marco Giannelli

Applied Sciences · 28.01.2026 · 16 (3) · 1322

DOI: 10.3390/app16031322 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Applied Sciences, online erschienen am 28.01.2026. Verfasst von Patrizio Barca und 5 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 64 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This phantom study presents a thorough characterization of the physical image quality of a clinical whole-body photon-counting computed tomography (PCCT) scanner. Multiple quality metrics—noise, noise power spectrum (NPS), task transfer function (TTF), and detectability index (d′)—were analyzed across a range of reconstruction algorithms (filtered back projection, FBP, and Quantum Iterative Reconstruction, QIR, with strength levels Q1–Q4), and varying reconstruction kernels (Br40/Br60/Br76/Br98). Both standard (STD, 0.4 mm slice thickness) and high-resolution (HR, 0.2 mm slice thickness) reconstruction modes were assessed. QIR significantly reduced image noise (60–95%) compared to FBP, particularly with sharper kernels. Spatial resolution improved with increasing QIR strength level for smoother kernels and was further enhanced using HR mode with sharp kernels. HR mode exhibited better noise performance than STD with sharper reconstructions, due to the small pixel effect. While STD mode showed higher d′ values for larger objects, HR mode outperformed it for smaller objects and sharper kernels. Compared to a conventional energy-integrating computed tomography system, the PCCT scanner showed superior d′ values under similar settings. Overall, this study highlights the complex interplay between acquisition and reconstruction parameters on image quality, confirms the potential of PCCT technology, and underscores the need for further clinical validation.

Zitieren

APA

Barca, P., Masturzo, L., Masi, L. D., Traino, A., Cademartiri, F., & Giannelli, M. (2026). Characterization of the Physical Image Quality of a Clinical Photon-Counting Computed Tomography Scanner Across Multiple Acquisition and Reconstruction Settings. Applied Sciences, 16(3), 1322. https://doi.org/10.3390/app16031322

Vancouver

Barca P, Masturzo L, Masi LD, Traino A, Cademartiri F, Giannelli M. Characterization of the Physical Image Quality of a Clinical Photon-Counting Computed Tomography Scanner Across Multiple Acquisition and Reconstruction Settings. Applied Sciences. 2026;16(3):1322. doi:10.3390/app16031322

BibTeX
@article{Barca2026,
  author = {Barca, Patrizio and Masturzo, Luigi and Masi, Luca De and Traino, Antonio and Cademartiri, Filippo and Giannelli, Marco},
  title = {Characterization of the Physical Image Quality of a Clinical Photon-Counting Computed Tomography Scanner Across Multiple Acquisition and Reconstruction Settings},
  journal = {Applied Sciences},
  year = {2026},
  volume = {16},
  number = {3},
  pages = {1322},
  doi = {10.3390/app16031322}
}
Kostenlos starten →