A Performance Evaluation of the Vitek®2 AST-N440 Card for Colistin Susceptibility Testing of Carbapenem-Resistant Acinetobacter baumannii Complex Isolates Using Broth Microdilution as the Reference Method
Antibiotics · 16.04.2026 · 15 (4) · 404
DOI: 10.3390/antibiotics15040404 ↗Kurz gefasst
Fachartikel in der Zeitschrift Antibiotics, online erschienen am 16.04.2026. Verfasst von Dimitra Petropoulou und 8 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Microbiology, General Panarcadian Hospital of Tripolis Evangelistria, 22100 Tripolis, Greece. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Objectives: Accurate determination of colistin (COL) in vitro activity against carbapenem-resistant Acinetobacter baumannii complex (CRAB) isolates remains challenging, as the reference broth microdilution (BMD) method is labor-intensive and not routinely implemented in most clinical laboratories. Semi-automated susceptibility methods for colistin in the clinical laboratory require validation. The present study evaluated the performance characteristics of the recently introduced Vitek®2 card AST-N440 for COL antimicrobial susceptibility testing (AST) on CRAB isolates compared with a BMD-based reference method (ComASP Colistin).
Methods
A total of 176 single-patient CRAB isolates from two distinct tertiary Greek hospitals between 2024 and 2025 were included. COL susceptibility testing was performed using Vitek®2 AST-N440 and compared with BMD. Minimum inhibitory concentrations (MICs) were interpreted according to EUCAST breakpoints. Method performance was evaluated by calculating categorical (CA) and essential agreement (EA), sensitivity, specificity, positive and negative predictive values (PPV/NPV), and major (ME) and very major error rates (VME) according to ISO 20776-2.
Results
Compared with BMD, AST-N440 showed a sensitivity of 89.6% and a specificity of 62.3%, with a PPV and NPV of 81.7% and 76.0%, respectively. The CA (80.1%) and the EA (46.0%) were below ISO acceptance criteria. The VME rate was 10.4%, and the ME rate 37.7%. Identical MIC values were observed in 25.0% of the isolates, while Vitek®2 reported lower and higher MIC values than BMD in 46.6% and 28.4% of isolates, respectively.
Conclusions
The Vitek®2 AST-N440 card performed suboptimally for COL susceptibility testing in CRAB isolates. Further validation of automated systems for COL AST is needed.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Laboratory of Microbiology, General Panarcadian Hospital of Tripolis Evangelistria, 22100 Tripolis, Greece.
- Laboratory of Clinical Microbiology, School of Medicine, Attikon University Hospital, National and Kapodistrian University of Athens, 12462 Athens, Greece.
ORCID-Profile: Dimitra Petropoulou, Anastasios Ioannidis, Georgia Petropoulou, Polyxeni Karakosta, Georgios Alexandros Baziotis, Spyros Pournaras
Zitieren
APA
Petropoulou, D., Ioannidis, A., Kaminioti, C., Mparka, C., Mitropoulou, E., Petropoulou, G., Karakosta, P., Baziotis, G. A., & Pournaras, S. (2026). A Performance Evaluation of the Vitek®2 AST-N440 Card for Colistin Susceptibility Testing of Carbapenem-Resistant Acinetobacter baumannii Complex Isolates Using Broth Microdilution as the Reference Method. Antibiotics, 15(4), 404. https://doi.org/10.3390/antibiotics15040404
Vancouver
Petropoulou D, Ioannidis A, Kaminioti C, Mparka C, Mitropoulou E, Petropoulou G, et al. A Performance Evaluation of the Vitek®2 AST-N440 Card for Colistin Susceptibility Testing of Carbapenem-Resistant Acinetobacter baumannii Complex Isolates Using Broth Microdilution as the Reference Method. Antibiotics. 2026;15(4):404. doi:10.3390/antibiotics15040404
BibTeX
@article{Petropoulou2026,
author = {Petropoulou, Dimitra and Ioannidis, Anastasios and Kaminioti, Christina and Mparka, Christina and Mitropoulou, Evgenia and Petropoulou, Georgia and Karakosta, Polyxeni and Baziotis, Georgios Alexandros and Pournaras, Spyros},
title = {A Performance Evaluation of the Vitek®2 AST-N440 Card for Colistin Susceptibility Testing of Carbapenem-Resistant Acinetobacter baumannii Complex Isolates Using Broth Microdilution as the Reference Method},
journal = {Antibiotics},
year = {2026},
volume = {15},
number = {4},
pages = {404},
doi = {10.3390/antibiotics15040404}
}