Analysis of Intensity in Different Light Cure Units Used in Dentistry

Jameel MA Sulaiman

Al-Rafidain Dental Journal · 01.01.2010 · 10 (1) · 192-197

DOI: 10.33899/rden.2010.8981 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Al-Rafidain Dental Journal, online erschienen am 01.01.2010. Verfasst von Jameel MA Sulaiman. Die Arbeit wurde bisher 2-mal zitiert (Crossref).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

0.05) were detected between Intensity and vary-ing time at 5, 10, 20, 30 sec. Conclusions: The intensity of the curing light was strongly affected by the angles and distance. The decrease in light intensity was significant (p< 0.05) when the light tip placed at an angle 70 and a way to the aperture of the a curing radiometer, but the magnitude of the intensity it becomes more with the same procedure when the angle was perpendicular 900. The decrease in the light intensity of the light curing units, agree with the inverse square law for the distances 0 to 16 mm. The study found that there was no significant difference (p> 0.05) in light intensity between the light curing tip and a curing radiometer varying with time. This study notice that the intensity of the light cure type LED is less than QTH(Quartz Tungsten Halogen).]]>

Zitieren

APA

Sulaiman, J. M. (2010). Analysis of Intensity in Different Light Cure Units Used in Dentistry. Al-Rafidain Dental Journal, 10(1), 192-197. https://doi.org/10.33899/rden.2010.8981

Vancouver

Sulaiman JM. Analysis of Intensity in Different Light Cure Units Used in Dentistry. Al-Rafidain Dental Journal. 2010;10(1):192-197. doi:10.33899/rden.2010.8981

BibTeX
@article{Sulaiman2010,
  author = {Sulaiman, Jameel MA},
  title = {Analysis of Intensity in Different Light Cure Units Used in Dentistry},
  journal = {Al-Rafidain Dental Journal},
  year = {2010},
  volume = {10},
  number = {1},
  pages = {192--197},
  doi = {10.33899/rden.2010.8981}
}
Kostenlos starten →