Evaluation of femoral head bone quality by Hounsfield units: A predictor of implant failure for intertrochanteric fractures after intramedullary nail fixation
Frontiers in Surgery · 06.01.2023 · 9
DOI: 10.3389/fsurg.2022.816742 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Surgery, online erschienen am 06.01.2023. Verfasst von Jixing Fan und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedics, Peking University Third Hospital, Beijing, China. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Key R&D Program of China, National Natural Science Foundation of China, Peking University Third Hospital.
Abstract (Originalsprache)
Purpose
The aim of present study is to evaluate the femoral head bone quality by Hounsfield units and its relationship to the occurrence of implant failure for intertrochanteric fractures after intramedullary nail fixation.
Methods
This retrospective study assessed 160 intertrochanteric fractures treated with intramedullary fixation. Patients with and without implant failure were divided into failure and control groups, respectively. The demographic information, femoral head Hounsfield unit (HU) value, the reduction quality, status of posteromedial support and position of the screw/blade were collected and compared. The logistic regression analyses were performed to evaluate risk factors of implant failure in intertrochanteric fractures after intramedullary nail fixation.
Results
Of the patients, 15 (9.38%) suffered from implant failure after intramedullary fixation. The mean HU value of femoral head was much lower in the failure group than the control group (133.25 ± 34.10 vs. 166.12 ± 42.68, p = 0.004). And the univariate analyses showed that A3 fracture and poor reduction quality were associated with implant failure ( p < 0.05). After adjustment for confounding variables, the multivariable logistic regression analyzes showed that femoral head HU value (odds ratio [OR], 0.972; 95% CI, 0.952–0.993; p = 0.008) and poor reduction quality (OR, 7.614; 95% CI, 1.390–41.717; p = 0.019) were independent influencing factors for implant failure.
Conclusion
The femoral head HU value was significantly correlated with the incidence of implant failure and can be used as an independent factor to predict implant failure for intertrochanteric fractures after intramedullary fixation.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopedics, Peking University Third Hospital, Beijing, China.
- Engineering Research Center of Bone and Joint Precision Medicine, Ministry of Education, Beijing, China.
Förderung
- National Key R&D Program of China
- National Natural Science Foundation of China
- Peking University Third Hospital
Zitieren
APA
Fan, J., Lv, Y., Xu, X., Zhou, F., Zhang, Z., Tian, Y., Ji, H., Guo, Y., Yang, Z., & Hou, G. (2023). Evaluation of femoral head bone quality by Hounsfield units: A predictor of implant failure for intertrochanteric fractures after intramedullary nail fixation. Frontiers in Surgery, 9. https://doi.org/10.3389/fsurg.2022.816742
Vancouver
Fan J, Lv Y, Xu X, Zhou F, Zhang Z, Tian Y, et al. Evaluation of femoral head bone quality by Hounsfield units: A predictor of implant failure for intertrochanteric fractures after intramedullary nail fixation. Frontiers in Surgery. 2023;9. doi:10.3389/fsurg.2022.816742
BibTeX
@article{Fan2023,
author = {Fan, Jixing and Lv, Yang and Xu, Xiangyu and Zhou, Fang and Zhang, Zhishan and Tian, Yun and Ji, Hongquan and Guo, Yan and Yang, Zhongwei and Hou, Guojin},
title = {Evaluation of femoral head bone quality by Hounsfield units: A predictor of implant failure for intertrochanteric fractures after intramedullary nail fixation},
journal = {Frontiers in Surgery},
year = {2023},
volume = {9},
doi = {10.3389/fsurg.2022.816742}
}
EUVIMED