Optimization of secondary flow path clearance in centrifugal blood pump: a combined numerical and experimental study

Chenying Zhu, Ping Ye, Zhaohua Chang

Frontiers in Physiology · 27.06.2025 · 16

DOI: 10.3389/fphys.2025.1595588 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Physiology, online erschienen am 27.06.2025. Verfasst von Chenying Zhu und 2 weiteren Autor:innen. Beteiligte Einrichtung: School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai, China. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Blood pumps as mechanical circulatory support (MCS) devices are widely used for patients with ventricular heart failure, and improving blood compatibility remains a key focus for researchers.

Methods

The effects of secondary flow path clearance on the performance and hemocompatibility of a centrifugal blood pump are investigated in this study. A blood pump of the specific design was developed for modeling and computer simulation, and 25 models were generated by varying its internal and outer clearance between 600 and 1,000 μm. This study systematically investigates the hydrodynamic and hemodynamic performance of designed blood pumps through integrated computational fluid dynamics simulations and in vitro hydraulic experiments.

Results

Key findings reveal a strong agreement between numerical predictions and experimental data, showing a maximum pressure deviation of 9% for the baseline pump configuration. Furthermore, the effect of the change in clearance on the head and efficiency is less than ±3%. The change in the secondary flow path significantly impacts hemolysis performance.

Discussion

An optimal clearance combination exists in the designed model, and for a specific flow field structure, asymmetric clearance may be an effective means to balance leakage control and turbulence suppression. Although the findings are specific to the pumps examined in this study, they provide valuable insights into the optimal design of blood pumps.

Schlagwörter

Stichwörter der Autor:innen

centrifugal blood pumpfluid dynamicshemolysissecondary flow clearanceshear stress

Autor:innen und Institutionen

  • School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai, China.

Zitieren

APA

Zhu, C., Ye, P., & Chang, Z. (2025). Optimization of secondary flow path clearance in centrifugal blood pump: a combined numerical and experimental study. Frontiers in Physiology, 16. https://doi.org/10.3389/fphys.2025.1595588

Vancouver

Zhu C, Ye P, Chang Z. Optimization of secondary flow path clearance in centrifugal blood pump: a combined numerical and experimental study. Frontiers in Physiology. 2025;16. doi:10.3389/fphys.2025.1595588

BibTeX
@article{Zhu2025,
  author = {Zhu, Chenying and Ye, Ping and Chang, Zhaohua},
  title = {Optimization of secondary flow path clearance in centrifugal blood pump: a combined numerical and experimental study},
  journal = {Frontiers in Physiology},
  year = {2025},
  volume = {16},
  doi = {10.3389/fphys.2025.1595588}
}
Kostenlos starten →