Comparing body temperature measurements using the double sensor method within a wearable device with oral and core body temperature measurements using medical grade thermometers—a short report
Frontiers in Physiology · 14.11.2023 · 14
DOI: 10.3389/fphys.2023.1279314 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Physiology, online erschienen am 14.11.2023. Verfasst von Arik Eisenkraft und 7 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Research in Military Medicine, Faculty of Medicine, The Hebrew University of Jerusalem and the Israel Defense Force Medical Corps, Jeru…. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Body temperature is essential for diagnosing, managing, and following multiple medical conditions. There are several methods and devices to measure body temperature, but most do not allow continuous and prolonged measurement of body temperature. Noninvasive skin temperature sensor combined with a heat flux sensor, also known as the “double sensor” technique, is becoming a valuable and simple method for frequently monitoring body temperature.
Methods
Body temperature measurements using the “double sensor” method in a wearable monitoring device were compared with oral and core body temperature measurements using medical grade thermometers, analyzing data from two prospective clinical trials of different clinical scenarios. One study included 45 hospitalized COVID-19 patients in which oral measurements were taken using a hand-held device, and the second included 18 post-cardiac surgery patients in which rectal measurements were taken using a rectal probe.
Results
In study 1, Bland-Altman analysis showed a bias of −0.04°C [0.34–(−0.43)°C, 95% LOA] with a correlation of 99.4% ( p < 0.001). In study 2, Bland-Altman analysis showed a bias of 0.0°C [0.27–(−0.28)°C, 95% LOA], and the correlation was 99.3% ( p < 0.001). In both studies, stratifying patients based on BMI and skin tone showed high accordance in all sub-groups.
Discussion
The wearable monitor showed high correlation with oral and core body temperature measurements in different clinical scenarios.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute for Research in Military Medicine, Faculty of Medicine, The Hebrew University of Jerusalem and the Israel Defense Force Medical Corps, Jerusalem, Israel.
- Biobeat Technologies Ltd., Petah Tikva, Israel.
- Heart Institute, Hadassah Medical Center, The Hebrew University of Jerusalem, Jerusalem, Israel.
Zitieren
APA
Eisenkraft, A., Goldstein, N., Fons, M., Tabi, M., Sherman, A. D., Ben Ishay, A., Merin, R., & Nachman, D. (2023). Comparing body temperature measurements using the double sensor method within a wearable device with oral and core body temperature measurements using medical grade thermometers—a short report. Frontiers in Physiology, 14. https://doi.org/10.3389/fphys.2023.1279314
Vancouver
Eisenkraft A, Goldstein N, Fons M, Tabi M, Sherman AD, Ben Ishay A, et al. Comparing body temperature measurements using the double sensor method within a wearable device with oral and core body temperature measurements using medical grade thermometers—a short report. Frontiers in Physiology. 2023;14. doi:10.3389/fphys.2023.1279314
BibTeX
@article{Eisenkraft2023,
author = {Eisenkraft, Arik and Goldstein, Nir and Fons, Meir and Tabi, Michael and Sherman, Anna Danielle and Ben Ishay, Arik and Merin, Roei and Nachman, Dean},
title = {Comparing body temperature measurements using the double sensor method within a wearable device with oral and core body temperature measurements using medical grade thermometers—a short report},
journal = {Frontiers in Physiology},
year = {2023},
volume = {14},
doi = {10.3389/fphys.2023.1279314}
}
EUVIMED