Comparison of PET photon attenuation for human tissue compositions, tissue-equivalent material compositions, and CT-phantom measurements
Frontiers in Physics · 24.07.2026 · 14
DOI: 10.3389/fphy.2026.1819239 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Physics, online erschienen am 24.07.2026. Verfasst von Heba Alrakh und 10 weiteren Autor:innen. Die Arbeit nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
In this study, the accuracy of linear attenuation coefficients for annihilation photons (511 keV) and additional cascade gammas (603, 1,077, and 1,157 keV) derived from two commercially available phantoms, i.e., the Sun Nuclear electron density and DIAGNOMATIC Pro-RT ED phantoms, was evaluated.
Methods
LACs obtained from computations using the NIST XCOM database and literature values of real tissue compositions were compared to LACs obtained from computations using the NIST XCOM database and tissue-equivalent inserts (when available) and to LACs obtained from transmission measurements. Also, LACs obtained from computations using the NIST XCOM database and tissue-equivalent inserts were compared to LACs obtained from transmission measurements (when possible). In addition, transmission scans were performed using the ECAT EXACT HR + positron emission tomography scanner.
Results
Twelve out of 23 tissue-equivalent inserts showed a deviation in the LACs of less than 5% when the values obtained from transmission measurements were compared to those computed using the NIST XCOM database with literature values of real tissue compositions. The highest relative errors were observed in lung-equivalent tissues (up to 37.45) and solid dense bone inserts (26.10%).
Discussion
Combining the breast and solid trabecular bone inserts of the Diagnomatic Pro-RT ED phantom with the liver, brain, adipose, lung-450, inner bone, cortical bone, HE Blood 40, HE Blood 70 and HE Blood 100 inserts from the Sun Nuclear phantom provides a relatively complete set of tissue-equivalent inserts from CT phantoms that reproduce the attenuation with systematic errors below 5%.
Zitieren
APA
Alrakh, H., Tellmann, L., Nikoubashman, O., Zimmermann, M., Shokrgozar, S., Winz, O. H., Faby, S., arra, A. A., Wiesmann, M., Shah, N. J., & Lerche, C. (2026). Comparison of PET photon attenuation for human tissue compositions, tissue-equivalent material compositions, and CT-phantom measurements. Frontiers in Physics, 14. https://doi.org/10.3389/fphy.2026.1819239
Vancouver
Alrakh H, Tellmann L, Nikoubashman O, Zimmermann M, Shokrgozar S, Winz OH, et al. Comparison of PET photon attenuation for human tissue compositions, tissue-equivalent material compositions, and CT-phantom measurements. Frontiers in Physics. 2026;14. doi:10.3389/fphy.2026.1819239
BibTeX
@article{Alrakh2026,
author = {Alrakh, Heba and Tellmann, Lutz and Nikoubashman, Omid and Zimmermann, Maike and Shokrgozar, Sadaf and Winz, Oliver H. and Faby, Sebastian and arra, Ali Abu and Wiesmann, Martin and Shah, N. Jon and Lerche, Christoph},
title = {Comparison of PET photon attenuation for human tissue compositions, tissue-equivalent material compositions, and CT-phantom measurements},
journal = {Frontiers in Physics},
year = {2026},
volume = {14},
doi = {10.3389/fphy.2026.1819239}
}
EUVIMED