Highly accurate and reliable ultrasonic focusing capability in heterogeneous media using a spherical cavity transducer
Frontiers in Physics · 13.04.2023 · 11
DOI: 10.3389/fphy.2023.1135744 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Physics, online erschienen am 13.04.2023. Verfasst von Dan Song und 8 weiteren Autor:innen. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China.
Abstract (Originalsprache)
Introduction
Focused ultrasound ablation surgery (FUAS) has been emerging to treat a wide range of conditions non-invasively and effectively with promising therapeutic outcomes. The focusing capability of an ultrasound transducer (i.e., focus shift, beam distortion, and acoustic pressure at the focus) determines the ablation effects. However, the focus shift and focal beam distortion after ultrasound propagating through multi-layered heterogeneous viscoelastic biological tissues become significant and are found to deteriorate the performance of FUAS in clinics.
Methods
To achieve an accurate and reliable focal field among patients with large variations in the anatomical structures and properties, a spherical cavity transducer with open ends and sub-wavelength focal size (Li et al., APL, 2013,102:204102) was applied here. Both experimental measurements and numerical simulations were performed to characterize the acoustic fields of the spherical cavity transducer in water, the multi-layered concentric cylindrical phantom, and the heterogeneous tissue model (an adult male pelvis enclosed by porcine skin, fat, and muscle) and then compared with those of a conventional concave transducer at the same electrical power output.
Results
It is found that standing-wave focusing using the spherical cavity transducer results in much less focus shift (0.25λ vs . 1.67λ) along the transducer axis and focal beam distortion (−6 dB beam area of 0.71 mm 2 vs . 4.72 mm 2 in water and 2.55 mm 2 vs . 17.30 mm 2 in tissue) in the focal plane but higher pressure focusing gain (40.05 dB vs . 33.61 dB in tissue).
Discussion
Such a highly accurate and reliable focal field is due to the excitation at an appropriate eigen-frequency of the spherical cavity with the varied media inside rather than the reverberation from the concave surface. Together with its sub-wavelength focal size, the spherical cavity transducer is technically advantageous in comparison to the concave one. The improved focusing capability would benefit ultrasound exposure for not only safer and more effective FUAS in clinics, but also broad acoustic applications.
Förderung
- National Natural Science Foundation of China
Zitieren
APA
Song, D., Xiong, J., Luo, M., Gong, X., Huang, W., Zeng, C., Zhou, Y., Li, F., & Wang, Z. (2023). Highly accurate and reliable ultrasonic focusing capability in heterogeneous media using a spherical cavity transducer. Frontiers in Physics, 11. https://doi.org/10.3389/fphy.2023.1135744
Vancouver
Song D, Xiong J, Luo M, Gong X, Huang W, Zeng C, et al. Highly accurate and reliable ultrasonic focusing capability in heterogeneous media using a spherical cavity transducer. Frontiers in Physics. 2023;11. doi:10.3389/fphy.2023.1135744
BibTeX
@article{Song2023,
author = {Song, Dan and Xiong, Jiupeng and Luo, Man and Gong, Xiaobo and Huang, Wenjun and Zeng, Chen and Zhou, Yufeng and Li, Faqi and Wang, Zhibiao},
title = {Highly accurate and reliable ultrasonic focusing capability in heterogeneous media using a spherical cavity transducer},
journal = {Frontiers in Physics},
year = {2023},
volume = {11},
doi = {10.3389/fphy.2023.1135744}
}
EUVIMED