Impulse oscillometry values calibrated against spirometric obstruction in children with suspected asthma
Frontiers in Pediatrics · 08.06.2026 · 14
DOI: 10.3389/fped.2026.1746282 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Pediatrics, online erschienen am 08.06.2026. Verfasst von Liang-Mei Lin und 4 weiteren Autor:innen. Beteiligte Einrichtung: Pediatric Respiratory Therapy Unit, Changhua Christian Children's Hospital, Changhua, Taiwan. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Kinderkrankenpflege
Abstract (Originalsprache)
Background
The impulse oscillation system (IOS) requires less cooperation than spirometry for assessing obstructive lung disease in children aged 6–18 years; however, diagnostic criteria for several IOS parameters remain controversial. This study compared IOS and spirometry to determine IOS parameter cutoff values for spirometry-detectable airway obstruction in children.
Methods
Spirometry was used as the reference standard. We enrolled 722 children with suspected asthma. Airway obstruction was defined as a ≥ 12% increase in FEV₁ after bronchodilator or any abnormal spirometric value (FEV₁ < 80% predicted, FEV₁/FVC <80%, or FEF parameters (MMEF, FEF₂₅–₇₅, etc.) < 56% predicted. IOS parameters (R5, X5, AX, Fres and R5–R20) were recorded.
Results
Spirometry and IOS parameters were significantly correlated, R5−R20 correlated the strongest with FEV1 and MMEF. The suggestive AX and Fres cutoff values were 1.54 kPa·L⁻ 1 and 20.41 Hz for 6–11-year olds and 0.54 kPa·L⁻ 1 and 13.77 Hz for 12–18-year olds. After bronchodilator inhalation, Fres and AX decreased respectively by 20% and 40% in 6–11-year olds, predicting obstructive lung diseases with accuracies of 62.2% and 60.2%. In the 12–18-year group, a 20% decrease in Fres and a 40% decrease in AX predicted obstructive lung disease with accuracies of 73.2% and 75.7%, respectively.
Conclusion
Our study compared pre- and post-bronchodilator IOS and spirometry results and identified which IOS values correspond to spirometric obstruction.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Pediatric Respiratory Therapy Unit, Changhua Christian Children's Hospital, Changhua, Taiwan.
- Department of Pediatric Immunology, Changhua Christian Children's Hospital, Changhua, Taiwan.
- School of Medicine, Kaohsiung Medical University, Kaohsiung, Taiwan.
- School of Medicine, Chung Shan Medical University, Taichung, Taiwan.
- Big Data Center, Changhua Christian Hospital, Changhua, Taiwan.
- Department of Pediatric Pulmonology and Critical Care, Changhua Christian Children's Hospital, Changhua, Taiwan.
Zitieren
APA
Lin, L. M., Tsai, Y. G., Chang, Y. J., Whang, J. E., & Lee, M. S. (2026). Impulse oscillometry values calibrated against spirometric obstruction in children with suspected asthma. Frontiers in Pediatrics, 14. https://doi.org/10.3389/fped.2026.1746282
Vancouver
Lin LM, Tsai YG, Chang YJ, Whang JE, Lee MS. Impulse oscillometry values calibrated against spirometric obstruction in children with suspected asthma. Frontiers in Pediatrics. 2026;14. doi:10.3389/fped.2026.1746282
BibTeX
@article{Lin2026,
author = {Lin, Liang-Mei and Tsai, Yi-Giien and Chang, Yu-Jun and Whang, Jeffrey Eli and Lee, Ming-Sheng},
title = {Impulse oscillometry values calibrated against spirometric obstruction in children with suspected asthma},
journal = {Frontiers in Pediatrics},
year = {2026},
volume = {14},
doi = {10.3389/fped.2026.1746282}
}
EUVIMED