Impulse oscillometry values calibrated against spirometric obstruction in children with suspected asthma

Liang-Mei Lin, Yi-Giien Tsai, Yu-Jun Chang, Jeffrey Eli Whang, Ming-Sheng Lee

Frontiers in Pediatrics · 08.06.2026 · 14

DOI: 10.3389/fped.2026.1746282 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Pediatrics, online erschienen am 08.06.2026. Verfasst von Liang-Mei Lin und 4 weiteren Autor:innen. Beteiligte Einrichtung: Pediatric Respiratory Therapy Unit, Changhua Christian Children's Hospital, Changhua, Taiwan. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The impulse oscillation system (IOS) requires less cooperation than spirometry for assessing obstructive lung disease in children aged 6–18 years; however, diagnostic criteria for several IOS parameters remain controversial. This study compared IOS and spirometry to determine IOS parameter cutoff values for spirometry-detectable airway obstruction in children.

Methods

Spirometry was used as the reference standard. We enrolled 722 children with suspected asthma. Airway obstruction was defined as a ≥ 12% increase in FEV₁ after bronchodilator or any abnormal spirometric value (FEV₁ < 80% predicted, FEV₁/FVC <80%, or FEF parameters (MMEF, FEF₂₅–₇₅, etc.) < 56% predicted. IOS parameters (R5, X5, AX, Fres and R5–R20) were recorded.

Results

Spirometry and IOS parameters were significantly correlated, R5−R20 correlated the strongest with FEV1 and MMEF. The suggestive AX and Fres cutoff values were 1.54 kPa·L⁻ 1 and 20.41 Hz for 6–11-year olds and 0.54 kPa·L⁻ 1 and 13.77 Hz for 12–18-year olds. After bronchodilator inhalation, Fres and AX decreased respectively by 20% and 40% in 6–11-year olds, predicting obstructive lung diseases with accuracies of 62.2% and 60.2%. In the 12–18-year group, a 20% decrease in Fres and a 40% decrease in AX predicted obstructive lung disease with accuracies of 73.2% and 75.7%, respectively.

Conclusion

Our study compared pre- and post-bronchodilator IOS and spirometry results and identified which IOS values correspond to spirometric obstruction.

Schlagwörter

Stichwörter der Autor:innen

bronchodilator testimpulse oscillometryobstructive lung diseasepediatricsspirometry

Autor:innen und Institutionen

  • Pediatric Respiratory Therapy Unit, Changhua Christian Children's Hospital, Changhua, Taiwan.
  • Department of Pediatric Immunology, Changhua Christian Children's Hospital, Changhua, Taiwan.
  • School of Medicine, Kaohsiung Medical University, Kaohsiung, Taiwan.
  • School of Medicine, Chung Shan Medical University, Taichung, Taiwan.
  • Big Data Center, Changhua Christian Hospital, Changhua, Taiwan.
  • Department of Pediatric Pulmonology and Critical Care, Changhua Christian Children's Hospital, Changhua, Taiwan.

Zitieren

APA

Lin, L. M., Tsai, Y. G., Chang, Y. J., Whang, J. E., & Lee, M. S. (2026). Impulse oscillometry values calibrated against spirometric obstruction in children with suspected asthma. Frontiers in Pediatrics, 14. https://doi.org/10.3389/fped.2026.1746282

Vancouver

Lin LM, Tsai YG, Chang YJ, Whang JE, Lee MS. Impulse oscillometry values calibrated against spirometric obstruction in children with suspected asthma. Frontiers in Pediatrics. 2026;14. doi:10.3389/fped.2026.1746282

BibTeX
@article{Lin2026,
  author = {Lin, Liang-Mei and Tsai, Yi-Giien and Chang, Yu-Jun and Whang, Jeffrey Eli and Lee, Ming-Sheng},
  title = {Impulse oscillometry values calibrated against spirometric obstruction in children with suspected asthma},
  journal = {Frontiers in Pediatrics},
  year = {2026},
  volume = {14},
  doi = {10.3389/fped.2026.1746282}
}
Kostenlos starten →