EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Pocket-sized hand-held ultrasound for evaluating pediatric functional constipation by both novice and expert clinicians

Pola Waissman, Ron Berant, Lisa Amir, Shimon Reif, Shmuel Springer

Frontiers in Pediatrics · 14.08.2025 · 13

DOI: 10.3389/fped.2025.1624070 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Pediatrics, online erschienen am 14.08.2025. Verfasst von Pola Waissman und 4 weiteren Autor:innen. Beteiligte Einrichtung: Neuromuscular and Human Performance Laboratory, Department of Physiotherapy, Ariel University, Ariel, Israel. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background/Objectives Pediatric functional constipation (PFC) is a global health concern. This study evaluates the utility of pocket-sized ultrasound (PsUS) as a tool for assessing PFC among clinicians with varying levels of ultrasound (US) experience. We assessed the validity of PsUS in measuring transverse rectal diameter (TRD) and rectal anterior wall thickness (RAWT) compared to conventional US and to evaluate agreement between expert and novice clinicians.

Methods

In this cross-sectional study, TRD and RAWT were measured using conventional US and PsUS in 52 children (PFC: n = 28; non-constipated: n = 24), aged 4–14 years. Measurements were performed by an experienced and a novice clinician. Agreement and validity were assessed using intraclass correlation coefficients (ICCs), and diagnostic accuracy was evaluated using receiver operating characteristic (ROC) analysis.

Results

High agreement was found between the experienced and novice clinicians in assessing PFC using conventional US (ICC for TRD = 0.98, 95% CI: 0.98–0.99; ICC for RAWT = 0.98, 95% CI: 0.97–0.99) and PsUS (ICC for TRD = 0.99, 95% CI: 0.97–0.99; ICC for RAWT = 0.97, 95% CI: 0.96–0.98). PsUS showed robust validity compared to conventional US (ICCs of 0.96 for TRD and 0.87 for RAWT). ROC analysis showed high diagnostic accuracy for PsUS at TRD [area under the curve (AUC) = 0.88, cut-off 30.25 mm] and RAWT (AUC = 0.91, cut-off 2.05 mm).

Conclusions

These results suggest PsUS is a valid and reliable tool for assessing PFC, that can be used by clinicians with varying levels of US experience.

Schlagwörter

Stichwörter der Autor:innen

pediatric functional constipationpocket-sized ultrasoundpoint-of-care ultrasoundrectal wall thicknesstransverse rectal diameter

Autor:innen und Institutionen

  • Neuromuscular and Human Performance Laboratory, Department of Physiotherapy, Ariel University, Ariel, Israel.
  • Department of Pediatric Emergency Medicine, Schneider Childrens Medical Center of Israel, Petach Tikva, Israel.
  • Department of Pediatrics, Hadassah-Hebrew University Medical Center, Jerusalem, Israel.

Zitieren

APA

Waissman, P., Berant, R., Amir, L., Reif, S., & Springer, S. (2025). Pocket-sized hand-held ultrasound for evaluating pediatric functional constipation by both novice and expert clinicians. Frontiers in Pediatrics, 13. https://doi.org/10.3389/fped.2025.1624070

Vancouver

Waissman P, Berant R, Amir L, Reif S, Springer S. Pocket-sized hand-held ultrasound for evaluating pediatric functional constipation by both novice and expert clinicians. Frontiers in Pediatrics. 2025;13. doi:10.3389/fped.2025.1624070

BibTeX
@article{Waissman2025,
  author = {Waissman, Pola and Berant, Ron and Amir, Lisa and Reif, Shimon and Springer, Shmuel},
  title = {Pocket-sized hand-held ultrasound for evaluating pediatric functional constipation by both novice and expert clinicians},
  journal = {Frontiers in Pediatrics},
  year = {2025},
  volume = {13},
  doi = {10.3389/fped.2025.1624070}
}
Kostenlos starten →