Fachartikel in der Zeitschrift Frontiers in Pediatrics, online erschienen am 03.10.2024. Verfasst von Albert Fabregat-Sanjuan und 6 weiteren Autor:innen. Beteiligte Einrichtung: FUNCMAT, Mechanical Engineering Department, Universitat Rovira i Virgili, Tarragona, Spain. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Neonatologists and clinical neurophysiologists face challenges with the current electrodes used for long-duration amplitude-integrated electroencephalography (aEEG) in neonatal intensive care units (NICU), limiting the capacity to diagnose brain damage.
Objectives
The objectives of this study were to develop methods for comparing the performance of different electrodes to be used in aEEG. The comparison was done between a newly designed neonate-specific electrode, aCUP-E, with commercial liquid gel electrodes used in amplitude-integrated electroencephalography (aEEG). The comparison included impedance stability, electrode survival, recording quality, usability, and satisfaction of NICU staff. Methods aEEG recordings with bipolar montage was used, with one hemisphere fitted with commercial electrodes and the other with aCUP-E electrodes, alternated among subjects. Continuous impedance and raw EEG data were collected over a minimum of 24 h, and signal processing was performed using Python and MATLAB. Main results aCUP-E electrodes demonstrated superior performance, including: Increased impedance stability and electrode survival, enhanced recording quality with fewer artifacts, high correlation in signal capture between electrodes during optimal brain activity segments, higher signal-to-noise ratio (SNR) across varying impedance levels, greater staff satisfaction and ease of use. Moreover, Kaplan-Meier curves indicated a higher survival rate for aCUP-E electrodes over 24 h compared to commercial electrodes. Impedance variability analysis showed statistically significant stability improvements for aCUP-E. Conclusion aCUP-E electrodes outperform commercial liquid gel electrodes in impedance stability, electrode survival, and recording quality. These results suggest that aCUP-E electrodes could significantly enhance aEEG utilization in diagnosing and treating neonatal brain conditions in NICUs. Future improvements to the aCUP-E electrode may further reduce artifacts and increase electrode longevity, potentially leading to a significant improvement in neonatal brain monitoring by means of aEEG.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- FUNCMAT, Mechanical Engineering Department, Universitat Rovira i Virgili, Tarragona, Spain.
- NeuroÈpia, Institut d'Investigació Sanitària Pere Virgili, Clinical Neurophysiology Department, Hospital Sant Joan de Reus, Reus, Spain.
- BIOMEC, Mechanical Engineering Department, Universitat Politècnica de Catalunya, Barcelona, Spain.
Zitieren
APA
Fabregat-Sanjuan, A., Rodríguez-Ballabriga, Á., Rigo-Vidal, A., Pàmies-Vilà, R., Larrosa-Capaces, S., Rius-Costa, V., & Pascual-Rubio, V. (2024). Analysis of electrode performance on amplitude integrated electroencephalography in neonates: evaluation of a new electrode aCUP-E vs. liquid gel electrodes. Frontiers in Pediatrics, 12. https://doi.org/10.3389/fped.2024.1452862
Vancouver
Fabregat-Sanjuan A, Rodríguez-Ballabriga Á, Rigo-Vidal A, Pàmies-Vilà R, Larrosa-Capaces S, Rius-Costa V, et al. Analysis of electrode performance on amplitude integrated electroencephalography in neonates: evaluation of a new electrode aCUP-E vs. liquid gel electrodes. Frontiers in Pediatrics. 2024;12. doi:10.3389/fped.2024.1452862
BibTeX
@article{FabregatSanjuan2024,
author = {Fabregat-Sanjuan, Albert and Rodríguez-Ballabriga, Ángel and Rigo-Vidal, Agnès and Pàmies-Vilà, Rosa and Larrosa-Capaces, Susana and Rius-Costa, Vanesa and Pascual-Rubio, Vicenç},
title = {Analysis of electrode performance on amplitude integrated electroencephalography in neonates: evaluation of a new electrode aCUP-E vs. liquid gel electrodes},
journal = {Frontiers in Pediatrics},
year = {2024},
volume = {12},
doi = {10.3389/fped.2024.1452862}
}
EUVIMED