Dosimetric impact of titanium, zirconia, and customized subperiosteal dental implants in head and neck radiotherapy: an in vitro study

Kübra Özkaya Toraman, Serap Çatlı Dinç, Can Ilgın, Alp Saruhanoğlu, Özge Kuru Kavak, Levent Demirkol, Gökçe Uçar Alveroğlu, Hatice Bilge Becerir

Frontiers in Oncology · 22.07.2026 · 16

DOI: 10.3389/fonc.2026.1875805 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Oncology, online erschienen am 22.07.2026. Verfasst von Kübra Özkaya Toraman und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Oncology Institute, Istanbul University, Istanbul, Türkiye. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Bilimsel Araştirma Projeleri Birimi, Istanbul Üniversitesi.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

High-density dental implant materials may lead clinically relevant uncertainties in dose calculation and delivery during head and neck radiotherapy (RT). This study aimed to experimentally and computationally evaluate the dosimetric impact of titanium, zirconia, and customized subperiosteal implants. Material and methods Three implant types were positioned within artificial bone phantoms (Sawbones ® ) simulating a human mandible. For the subperiosteal group, a patient-specific framework was digitally designed and manufactured. Optically Stimulated Luminescence (OSL) dosimeters were placed at four anatomical locations (distal, buccal, mesial, and lingual) around each implant. Computed tomography (CT) scans (1-mm slice thickness) were obtained, and Intensity-Modulated Radiation Therapy (IMRT) plans were generated using a treatment planning system (TPS) equipped with the AcurosXB (AXB) algorithm with 6-MV photon energy and a target dose of 200 cGy. Statistical analysis was performed using the Wilcoxon Signed-Rank test on dependent groups to compare TPS-calculated and OSL-measured doses.

Results

A significant discrepancy was found between the overall measured and planned doses (p=0.003), with a median dose deviation of -7.8%. Backscatter-induced dose enhancement was systematically detected in the buccal region (up to +1.9%), whereas dose attenuation was predominant in the lingual and mesial regions (up to -14.5%). TPS-calculated maximum point doses revealed substantial hot spots, reaching 228 cGy for zirconia and 217 cGy for subperiosteal implants at the over-the-implant interface. While significant differences were confirmed for the endosseous group (p=0.021), the customized subperiosteal group demonstrated a trend toward a more predictable dosimetric profile (p=0.068) with a lower median deviation (-6.8%) compared to both titanium (-7.8%) and zirconia (-10.5%) endosseous implants.

Conclusions

Dental implant material and design influence dose distribution during RT, with high-density materials such as zirconia associated with greater discrepancies in dose estimation. Customized subperiosteal implants demonstrated a more predictable dosimetric profile in this experimental setting; however, the clinical significance of these findings remains to be established through larger studies incorporating clinical outcome data.

Schlagwörter

Stichwörter der Autor:innen

dental implantdosimetryhead-neckradiotherapysubperiosteal implanttitaniumtreatment planning systemzirconia

Autor:innen und Institutionen

  • Department of Radiation Oncology, Oncology Institute, Istanbul University, Istanbul, Türkiye.
  • Department of Radiation Oncology, Faculty of Medicine, Gazi University, Ankara, Türkiye.
  • Department of Radiation Oncology, Istanbul Faculty of Medicine, Istanbul University, Istanbul, Istanbul, Türkiye.
  • Department of Oral and Maxillofacial Surgery, Istanbul University Faculty of Dentistry, Istanbul, Türkiye.
  • Department of Medical Physics, Oncology Institute, Istanbul University, Istanbul, Türkiye.

Förderung

  • Bilimsel Araştirma Projeleri Birimi, Istanbul Üniversitesi

Zitieren

APA

Özkaya Toraman, K., Çatlı Dinç, S., Ilgın, C., Saruhanoğlu, A., Kuru Kavak, Ö., Demirkol, L., Uçar Alveroğlu, G., & Bilge Becerir, H. (2026). Dosimetric impact of titanium, zirconia, and customized subperiosteal dental implants in head and neck radiotherapy: an in vitro study. Frontiers in Oncology, 16. https://doi.org/10.3389/fonc.2026.1875805

Vancouver

Özkaya Toraman K, Çatlı Dinç S, Ilgın C, Saruhanoğlu A, Kuru Kavak Ö, Demirkol L, et al. Dosimetric impact of titanium, zirconia, and customized subperiosteal dental implants in head and neck radiotherapy: an in vitro study. Frontiers in Oncology. 2026;16. doi:10.3389/fonc.2026.1875805

BibTeX
@article{zkayaToraman2026,
  author = {Özkaya Toraman, Kübra and Çatlı Dinç, Serap and Ilgın, Can and Saruhanoğlu, Alp and Kuru Kavak, Özge and Demirkol, Levent and Uçar Alveroğlu, Gökçe and Bilge Becerir, Hatice},
  title = {Dosimetric impact of titanium, zirconia, and customized subperiosteal dental implants in head and neck radiotherapy: an in vitro study},
  journal = {Frontiers in Oncology},
  year = {2026},
  volume = {16},
  doi = {10.3389/fonc.2026.1875805}
}