A comparative dosimetric evaluation of two image-guided calibration strategies for multi-isocenter total body irradiation: integrated fan-beam CT–based alignment versus sequential cone-beam CT–guided localization

Bo Gao, Yaling Hong, Junjing Yan, Xiwei Chen, Suhua Tan, Yutao Zhu, Yan Liu, Zhaojie Yao, Lizhen Wu, Haiyan Chen, Zhanquan Lei

Frontiers in Oncology · 04.03.2026 · 16

DOI: 10.3389/fonc.2026.1760909 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Oncology, online erschienen am 04.03.2026. Verfasst von Bo Gao und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Fujian Children's Hospital (Fujian Branch of Shanghai Children's Medical Center), College of Clinical Medicine for …. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Science Foundation of Fujian Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Image-guided radiotherapy (IGRT) is key to making sure the radiation dose is accurate when treating kids with multi-isocenter total body irradiation (TBI). Ensuring dose accuracy in pediatric multi-isocenter TBI. Shanghai Children’s Medical Center employed low-dose fan-beam CT (FBCT) for integrated multi-isocenter calibration, whereas Fujian Children’s Hospital used cone-beam CT (CBCT) for sequential isocenter alignment.

Objective

To evaluate FBCT versus CBCT guidance effects on target coverage and organ doses across two regimens: 12 Gy in 6 fractions and 3 Gy in 1 fraction.

Methods

Retrospective analysis of 34 pediatric TBI patients (21 FBCT, 13 CBCT) treated with identical field setups (three upper-body isocenters; two/three lower-body isocenters; junction at upper femur third). Pre-treatment FBCT/CBCT images were registered to planning CTs; doses recalculated using original plans. Metrics: PTV V90%, V100%, V110%, mean doses; homogeneity index (HI); conformity index (CI); mean lung and kidney doses.

Results

In the 12 Gy group, FBCT guidance improved PTV coverage: V 90% increased from 96.11% to 97.14%, V 100% from 90.40% to 92.81%, and V 110% decreased from 20.73% to 16.67% (all P < 0.01). HI decreased from 0.25 to 0.16, CI increased from 0.77 to 0.89, and mean PTV dose rose from 12.33 to 12.57 Gy (all P < 0.01). Mean lung dose fell from 8.61 to 8.47 Gy, and mean kidney dose from 8.24 to 8.10 Gy (both P < 0.01). In the 3 Gy group, FBCT guidance also improved PTV coverage: V 90% increased from 96.32% to 97.82%, V 100% from 91.44% to 93.97%, and V 110% decreased from 17.43% to 13.72% (all P < 0.01). HI decreased from 0.21 to 0.13, CI increased from 0.77 to 0.87, and mean PTV dose rose from 3.08 to 3.12 Gy (all P < 0.01). Mean lung dose decreased from 2.34 to 2.25 Gy, and mean kidney dose from 2.09 to 2.06 Gy (both P < 0.01).

Conclusion

FBCT guidance gave better target dose conformity and homogeneity, and lower lung doses, than CBCT guidance—both for the 12 Gy myeloablative regimen and the 3 Gy low-dose regimen. These results suggest that FBCT guidance is a better option for image-guided total body irradiation in children.

Schlagwörter

Stichwörter der Autor:innen

childrencone-beam CTdosimetryfan-beam CTimage guidancetotal body irradiation

Autor:innen und Institutionen

  • Department of Radiation Oncology, Fujian Children's Hospital (Fujian Branch of Shanghai Children's Medical Center), College of Clinical Medicine for Obstetrics & Gynecology and Pediatrics, Fujian Medical University, Fuz…
  • Department of Radiation Oncology, Shanghai Children's Medical Center Affiliated to Shanghai Jiaotong University School of Medicine Institute, Shanghai, China.
  • Hunan Provincial Hospital of Integrated Traditional Chinese and Western Medicine, Changsha, Hunan, China.

Förderung

  • Natural Science Foundation of Fujian Province

Zitieren

APA

Gao, B., Hong, Y., Yan, J., Chen, X., Tan, S., Zhu, Y., Liu, Y., Yao, Z., Wu, L., Chen, H., & Lei, Z. (2026). A comparative dosimetric evaluation of two image-guided calibration strategies for multi-isocenter total body irradiation: integrated fan-beam CT–based alignment versus sequential cone-beam CT–guided localization. Frontiers in Oncology, 16. https://doi.org/10.3389/fonc.2026.1760909

Vancouver

Gao B, Hong Y, Yan J, Chen X, Tan S, Zhu Y, et al. A comparative dosimetric evaluation of two image-guided calibration strategies for multi-isocenter total body irradiation: integrated fan-beam CT–based alignment versus sequential cone-beam CT–guided localization. Frontiers in Oncology. 2026;16. doi:10.3389/fonc.2026.1760909

BibTeX
@article{Gao2026,
  author = {Gao, Bo and Hong, Yaling and Yan, Junjing and Chen, Xiwei and Tan, Suhua and Zhu, Yutao and Liu, Yan and Yao, Zhaojie and Wu, Lizhen and Chen, Haiyan and Lei, Zhanquan},
  title = {A comparative dosimetric evaluation of two image-guided calibration strategies for multi-isocenter total body irradiation: integrated fan-beam CT–based alignment versus sequential cone-beam CT–guided localization},
  journal = {Frontiers in Oncology},
  year = {2026},
  volume = {16},
  doi = {10.3389/fonc.2026.1760909}
}
Kostenlos starten →